The use of x-ray computed tomography for dimensional measurements associated with engineering applications has flourished in recent years. However, error sources associated with the technology are not well understood. In this paper, a novel two-sphere reference sample has been developed and used to investigate the stability of the imaging system that consists of an x-ray tube and a detector. In contrast with other research work reported, this work considered relative positional variation along the x-, y- and z-axes. This sample is a significant improvement over the one sphere sample proposed previously, which can only be used to observe the stability of the imaging system along x- and y-axes. Temperature variations of different parts of the system have been monitored and the relationship between temperature variations and x-ray image stability has been studied. Other effects that may also influence the stability of the imaging system have been discussed. The proposed reference sample and testing method are transferable to other types of x-ray computed tomography systems, for example, systems with transmission targets and systems with sub-micrometre focal spots.
机译:近年来,X射线计算机断层摄影技术在与工程应用相关的尺寸测量中的应用正在蓬勃发展。但是,与该技术相关的错误源尚未得到很好的理解。在本文中,已经开发了一种新颖的两球参考样品,并用于研究由X射线管和探测器组成的成像系统的稳定性。与报道的其他研究工作相反,该工作考虑了沿x,y和z轴的相对位置变化。该样品是对先前提出的单球样品的一项重大改进,该样品只能用于观察成像系统沿x轴和y轴的稳定性。监视了系统不同部分的温度变化,并研究了温度变化与X射线图像稳定性之间的关系。已经讨论了可能影响成像系统稳定性的其他影响。所提出的参考样品和测试方法可以转移到其他类型的X射线计算机断层摄影系统中,例如,具有透射目标的系统和具有亚微米焦点的系统。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号