An improvised Zirconia-Dolerite ball plate artefact is manufactured, which computes different geometric errors of CMM in three different plane positions (XY, YZ and ZX). Uncertainty measurement of this artefact will affect calibration of CMM. CMM undergoing periodic interim checks by ball plate measurement needs to compute uncertainty analysis over a calibration period. This requires repetitive measurements of artefact, involving great measurement cost. An attempt has been made in this research work to devise Monte Carlo framework (MCF) for analyzing the uncertainty of ball plate measurement. The presented MCF requires less repetitive measurement at actual. This work also made possible to find preferable zones of measuring areas of CMM including volumetric measurement, explaining novelty of the research work. In the present test specific CMM, demonstrated MCF recommends seldom use of lower left corner of CMM.
机译:制造了一种简易的氧化锆-白云石球板工件,该工件可在三个不同的平面位置(XY,YZ和ZX)计算CMM的不同几何误差。该伪像的不确定性测量将影响CMM的校准。通过球板测量进行定期中期检查的CMM需要在校准期内计算不确定性分析。这就需要对伪像进行重复测量,这需要大量的测量成本。在这项研究工作中已经尝试设计蒙特卡洛框架(MCF)来分析球板测量的不确定性。提出的MCF实际上需要较少的重复测量。这项工作还可以找到包括体积测量在内的CMM测量区域的最佳区域,从而说明了研究工作的新颖性。在当前测试专用的CMM中,演示的MCF建议很少使用CMM的左下角。