首页> 外文期刊>Microchemical Journal: Devoted to the Application of Microtechniques in all Branches of Science >On-line preconcentration and GFAAS determination of trace metals in waters
【24h】

On-line preconcentration and GFAAS determination of trace metals in waters

机译:在线预浓缩和GFAAS测定水中的痕量金属

获取原文
获取原文并翻译 | 示例
       

摘要

An on-line separation preconcentration system coupled to electrothermal (graphite furnace) atomic absorption spectrometry was developed. A miniature column packed with iminodiacetic acid ethyl cellulose (IDAEC) was inserted into the loop. A peristaltic pump was used to deliver solutions. A flow of air was driven into the packed column, evacuating it between sample loading, washing and elution. The retained analyte was introduced on-line to graphite furnace using countercurrent elution with HNO3. The system was applied for the determination of V, Co and Pb in medicinal mineral water samples, and nickel in sea water samples. The detection limits (3 sigma) were 0.058, 0.022, 0.067, 0.062 mug/l for Co, Ph, V, and Ni, respectively. The R.S.D. (n = 5) was < 5% at 0.4-1.0 g/l concentration range. (C) 2000 Elsevier Science B.V. All rights reserved. [References: 6]
机译:开发了与电热(石墨炉)原子吸收光谱法耦合的在线分离预浓缩系统。将装有亚氨基二乙酸乙基纤维素(IDAEC)的微型色谱柱插入回路中。使用蠕动泵来输送溶液。气流进入填充柱,在样品上样,洗涤和洗脱之间抽空。使用HNO3逆流洗脱将保留的分析物在线引入石墨炉。该系统用于测定药用矿泉水样品中的V,Co和Pb,以及海水样品中的镍。 Co,Ph,V和Ni的检出限(3 sigma)分别为0.058、0.022、0.067、0.062马克杯/升。 R.S.D. (n = 5)在0.4-1.0μg/ l的浓度范围内<5%。 (C)2000 Elsevier Science B.V.保留所有权利。 [参考:6]