摘要

There are currently nine caesium fountain primary frequency standards regularly reporting calibrations of International Atomic Time to the Bureau International des Poids et Measures (BIPM). An investigation has been carried out using data from the BIPM publication Circular T to evaluate the frequency differences among these standards and to determine whether these offsets are consistent with the stated uncertainties. The fractional frequency uncertainties of some Cs fountains are now in the range of 4 X 10~(-16) to 5 X 10~(-16). The results of this investigation show that the standards agree well with each other. An overall estimate of the caesium frequency is made using the weighted mean of all the fountains.
机译:当前,有九种铯源泉一次频率标准定期向国际仿制与测量局(BIPM)报告国际原子时间的校准。已使用BIPM出版物T通告中的数据进行了调查,以评估这些标准之间的频率差异,并确定这些偏移量是否与规定的不确定性一致。现在,某些Cs喷泉的分数频率不确定度在4 X 10〜(-16)到5 X 10〜(-16)的范围内。调查结果表明,这些标准彼此之间非常吻合。使用所有喷泉的加权平均值对铯频率进行总体估算。