Calibrated capsule-type standard platinum resistance thermometers were used to compare national realizations of the International Temperature Scale of 1990 (ITS-90) from 13.8033 K, the triple point of equilibrium hydrogen, to 273.16 K, the triple point of water, for seven countries in CIPM Key Comparison CCT-K2. Measurements were made at temperatures close to the eight low-temperature defining fixed points of the ITS-90, using a copper comparison block capable of simultaneously holding nine thermometers. Two separate measurement runs were performed, allowing two different groups of capsules from each laboratory to be examined. The results are used to determine the degree of equivalence of the independent national realizations of the scale for use in the Mutual Recognition Arrangement Appendix B database. In addition, measurements were made with the first group of thermometers at approximately eighty temperatures throughout the cryogenic range, which provide information to evaluate some of the so-called scale non-uniqueness issues inherent in the ITS-90 interpolation scheme.
机译:使用校准的胶囊型标准铂电阻温度计比较了七个国家在1990年国际温度标度(ITS-90)上从平衡氢的三相点13.8033 K到水的三相点273.16 K的国家实现情况。在CIPM密钥比较CCT-K2中。使用能够同时容纳九个温度计的铜比较块,在接近ITS-90的八个低温定义固定点的温度下进行测量。进行了两次单独的测量运行,从而可以检查来自每个实验室的两组不同的胶囊。结果用于确定相互认可安排附录B数据库中使用的比额表的独立国家实现的等效程度。此外,第一组温度计在整个低温范围内约80个温度下进行了测量,这些信息可提供信息以评估ITS-90内插方案中固有的一些所谓的比例尺非唯一性问题。