Smart instrumentation for determination of ligament stiffness and ligament balance in total knee arthroplasty

机译:用于确定全膝关节置换术中韧带刚度和韧带平衡的智能仪器

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Ligament balance is an important and subjective task performed during total knee arthroplasty (TKA) procedure. For this reason, it is desirable to develop instruments to quantitatively assess the soft-tissue balance since excessive imbalance can accelerate prosthesis wear and lead to early surgical revision. The instrumented distractor proposed in this study can assist surgeons on performing ligament balance by measuring the distraction gap and applied load. Also the device allows the determination of the ligament stiffness which can contribute a better understanding of the intrinsic mechanical behavior of the knee joint. Instrumentation of the device involved the use of hall-sensors for measuring the distractor displacement and strain gauges to transduce the force. The sensors were calibrated and tested to demonstrate their suitability for surgical use. Results show the distraction gap can be measured reliably with 0.1. mm accuracy and the distractive loads could be assessed with an accuracy in the range of 4. N. These characteristics are consistent with those have been proposed, in this work, for a device that could assist on performing ligament balance while permitting surgeons evaluation based on his experience. Preliminary results from in vitro tests were in accordance with expected stiffness values for medial collateral ligament (MCL) and lateral collateral ligament (LCL).
机译:韧带平衡是全膝关节置换术(TKA)手术中执行的重要主观任务。因此,由于过度的不平衡会加速假体的磨损并导致早期的外科手术翻修,因此需要开发一种能够定量评估软组织平衡的仪器。本研究中提出的器械式牵张器可以通过测量牵张间隙和施加的负荷来帮助外科医生进行韧带平衡。该装置还允许确定韧带刚度,这可以有助于更好地理解膝关节的固有机械行为。该设备的仪器包括使用霍尔传感器测量撑开器的位移以及使用应变仪来传递力。传感器经过校准和测试,以证明其适用于外科手术。结果表明分散间隙可以可靠地用0.1测量。 mm精度和分散负荷的精度可以在4 N范围内评估。这些特性与本工作中提出的那些特性相一致,该设备可以帮助进行韧带平衡,同时允许外科医生根据以下情况进行评估他的经验。体外测试的初步结果与内侧副韧带(MCL)和外侧副韧带(LCL)的预期刚度值一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号