...
首页> 外文期刊>Medical Physics >Verification of multileaf collimator leaf positions using an electronic portal imaging device.
【24h】

Verification of multileaf collimator leaf positions using an electronic portal imaging device.

机译:使用电子门户成像设备验证多叶准直仪叶片位置。

获取原文
获取原文并翻译 | 示例
           

摘要

An automated method is presented for determining individual leaf positions of the Siemens dual focus multileaf collimator (MLC) using the Siemens BEAMVIEW(PLUS) electronic portal imaging device (EPID). Leaf positions are computed with an error of 0.6 mm at one standard deviation (sigma) using separate computations of pixel dimensions, image distortion, and radiation center. The pixel dimensions are calculated by superimposing the film image of a graticule with the corresponding EPID image. A spatial correction is used to compensate for the optical distortions of the EPID, reducing the mean distortion from 3.5 pixels (uncorrected) per localized x-ray marker to 2 pixels (1 mm) for a rigid rotation and 1 pixel for a third degree polynomial warp. A correction for a nonuniform dosimetric response across the field of view of the EPID images is not necessary due to the sharp intensity gradients across leaf edges. The radiation center, calculated from the average of the geometric centers of a square field at 0degrees and 180 degrees collimator angles, is independent of graticule placement error. Its measured location on the EPID image was stable to within 1 pixel based on 3 weeks of repeated extensions/retractions of the EPID. The MLC leaf positions determined from the EPID images agreed to within a pixel of the corresponding values measured using film and ionization chamber. Several edge detection algorithms were tested: contour, Sobel, Roberts, Prewitt, Laplace, morphological, and Canny. These agreed with each other to within < or = 1.2 pixels for the in-air EPID images. Using a test pattern, individual MLC leaves were found to be typically within 1 mm of the corresponding record-and-verify values, with a maximum difference of 1.8 mm, and standard deviations of <0.3 mm in the daily reproducibility. This method presents a fast, automatic, and accurate alternative to using film or a light field for the verification and calibration of the MLC.
机译:提出了一种自动方法,该方法使用西门子BEAMVIEW(PLUS)电子门成像设备(EPID)确定西门子双焦点多叶准直仪(MLC)的各个叶片位置。使用像素尺寸,图像失真和辐射中心的单独计算,以一个标准偏差(sigma)的0.6 mm误差计算叶片位置。通过将标线的胶片图像与相应的EPID图像叠加来计算像素尺寸。使用空间校正来补偿EPID的光学失真,从而将平均失真从每个局部X射线标记的3.5像素(未校正)减少到刚性旋转的2像素(1毫米)和三次多项式的1像素经。由于跨叶边缘的强度梯度很陡,因此不需要跨EPID图像的视野对不均匀的剂量响应进行校正。辐射中心是根据正方形场在0度和180度准直器角度处的几何中心的平均值计算得出的,与网格放置误差无关。根据EPID重复延伸/缩回3周,其在EPID图像上的测量位置稳定在1个像素以内。根据EPID图像确定的MLC叶片位置在使用胶片和电离室测量的相应值的像素内一致。测试了几种边缘检测算法:轮廓,Sobel,Roberts,Prewitt,Laplace,morphological和Canny。空中EPID图像的这些像素彼此一致,误差在<或= 1.2像素以内。使用测试图案,发现单个MLC叶片通常在相应记录和验证值的1毫米之内,最大差异为1.8毫米,日再现性的标准偏差为<0.3毫米。这种方法提供了一种快速,自动和准确的替代方法,可替代使用胶片或光场进行MLC的验证和校准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号