首页> 外文期刊>Medical Physics >Uncertainties in effective dose estimates of adult CT head scans: the effect of head size.
【24h】

Uncertainties in effective dose estimates of adult CT head scans: the effect of head size.

机译:成人CT头部扫描的有效剂量估算不确定性:头部尺寸的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: This study is an extension of a previous study where the uncertainties in effective dose estimates from adult CT head scans were calculated using four CT effective dose estimation methods, three of which were computer programs (CT-EXPO, CTDOSIMETRY, and IMPACTDOSE) and one that involved the dose length product (DLP). However, that study did not include the uncertainty contribution due to variations in head sizes. METHODS: The uncertainties due to head size variations were estimated by first using the computer program data to calculate doses to small and large heads. These doses were then compared with doses calculated for the phantom heads used by the computer programs. An uncertainty was then assigned based on the difference between the small and large head doses and the doses of the phantom heads. RESULTS: The uncertainties due to head size variations alone were found to be between 4% and 26% depending on the method used and the patient gender. When these uncertainties were included with the results of the previous study, the overall uncertainties in effective dose estimates (stated at the 95% confidence interval) were 20%-31% (CT-EXPO), 15%-30% (CTDOSIMETRY), 20%-36% (IMPACTDOSE), and 31%-40% (DLP). CONCLUSIONS: For the computer programs, the lower overall uncertainties were still achieved when measured values of CT dose index were used rather than tabulated values. For DLP dose estimates, head size variations made the largest (for males) and second largest (for females) contributions to effective dose uncertainty. An improvement in the uncertainty of the DLP method dose estimates will be achieved if head size variation can be taken into account.
机译:目的:本研究是先前研究的扩展,该研究使用四种CT有效剂量估算方法计算了成人CT头部扫描的有效剂量估算的不确定性,其中三种是计算机程序(CT-EXPO,CTDOSIMETRY和IMPACTDOSE),以及一种涉及剂量长度乘积(DLP)的药物。但是,该研究未包括由于头部尺寸的变化而导致的不确定性贡献。方法:首先通过使用计算机程序数据来计算针对大大小小的头部的剂量,来估计由于头部大小变化而引起的不确定性。然后将这些剂量与为计算机程序使用的幻像头计算的剂量进行比较。然后根据小头和大头剂量与幻影头剂量之间的差异分配不确定性。结果:根据所用方法和患者性别的不同,仅因头部大小变化而导致的不确定性在4%至26%之间。当这些不确定性包括在先前研究的结果中时,有效剂量估算(以95%置信区间表示)的总体不确定性为20%-31%(CT-EXPO),15%-30%(CTDOSIMETRY), 20%-36%(IMPACTDOSE)和31%-40%(DLP)。结论对于计算机程序,当使用CT剂量指数的测量值而不是列表值时,总体不确定性仍然较低。对于DLP剂量估计,头部大小变化对有效剂量不确定性的影响最大(男性),第二大(女性)。如果可以考虑头部尺寸的变化,将可以改善DLP方法剂量估计的不确定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号