Sensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parameters.

机译:兆电压光子束的蒙特卡罗模拟对电子束和其他参数的敏感性。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The BEAM code is used to simulate nine photon beams from three major manufacturers of medical linear accelerators (Varian, Elekta, and Siemens), to derive and evaluate estimates for the parameters of the electron beam incident on the target, and to study the effects of some mechanical parameters like target width, primary collimator opening, flattening filter material and density. The mean energy and the FWHM of the incident electron beam intensity distributions (assumed Gaussian and cylindrically symmetric) are derived by matching calculated percentage depth-dose curves past the depth of maximum dose (within 1% of maximum dose) and off-axis factors (within 2sigma at 1% statistics or less) with measured data from the AAPM RTC TG-46 compilation. The off-axis factors are found to be very sensitive to the mean energy of the electron beam, the FWHM of its intensity distribution, its angle of incidence, the dimensions of the upper opening of the primary collimator, the material of the flattening filter and its density. The off-axis factors are relatively insensitive to the FWHM of the electron beam energy distribution, its divergence and the lateral dimensions of the target. The depth-dose curves are sensitive to the electron beam energy, and to its energy distribution, but they show no sensitivity to the FWHM of the electron beam intensity distribution. The electron beam incident energy can be estimated within 0.2 MeV when matching either the measured off-axis factors or the central-axis depth-dose curves when the calculated uncertainties are about 0.7% at the 1 sigma level. The derived FWHM (+/-0.1 mm) of the electron beam intensity distributions all fall within 1 mm of the manufacturer specifications except in one case where the difference is 1.2 mm.
机译:BEAM代码用于模拟来自医疗线性加速器的三个主要制造商(Varian,Elekta和Siemens)的九个光子束,以导出和评估入射在目标上的电子束参数的估计值,并研究电子束的影响。一些机械参数,例如目标宽度,主准直仪开口,滤光片材料的平整度和密度。通过匹配经过计算的最大剂量深度(最大剂量的1%以内)内的深度剂量曲线的百分比和离轴因子,得出入射电子束强度分布的平均能量和FWHM(假定为高斯分布和圆柱对称)。 AAPM RTC TG-46汇编中的测量数据在2sigma之内(统计值的1%或更少)。已发现离轴因子对电子束的平均能量,其强度分布的FWHM,其入射角,主准直仪上开口的尺寸,平坦滤光片的材料和它的密度。离轴因素对电子束能量分布的FWHM,其发散度和目标的横向尺寸相对不敏感。深度-剂量曲线对电子束能量及其能量分布敏感,但对电子束强度分布的FWHM则不敏感。当匹配的测得的离轴因数或中心轴深度-剂量曲线匹配时,当计算的不确定度在1 sigma时约为0.7%时,电子束入射能量可以估计在0.2 MeV以内。所导出的电子束强度分布的FWHM(+/- 0.1 mm)均落在制造商规定的1 mm之内,除非一种情况下的差为1.2 mm。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号