A formalism for an independent dose verification of the Gamma Knife treatment planning is developed. It is based on the approximation that isodose distribution for a single shot is in the shape of an ellipsoid in three-dimensional space. The dose profiles for a phantom along each of the three major axes are fitted to a function which contains the terms that represent the contributions from a point source, an extrafocal scattering, and a flat background. The fitting parameters are extracted for all four helmet collimators, at various shot locations, and with different skull shapes. The 33 parameters of a patient's skull shape obtained from the Skull Scaling Instrument measurements are modeled for individual patients. The relative doses for a treatment volume in the form of 31 x 31 x 31 matrix of points are extracted from the treatment planning system, the Leksell Gamma-Plan (LGP). Our model evaluates the relative doses using the same input parameters as in the LGP, which are skull measurement data, shot location, weight, gamma-angle of the head frame, and helmet collimator size. For 29 single-shot cases, the discrepancy of dose at the focus point between the calculation and the LGP is found to be within -1% to 2%. For multi-shot cases, the value and the coordinate of the maximum dose point from the calculation agree within +/-7% and +/-3 mm with the LGP results. In general, the calculated doses agree with the LGP calculations within +/-10% for the off-center locations. Results of calculation with this method for the dimension and location of the 50% isodose line are in good agreement with results from Leksell GammaPlan. Therefore, this method can be served as a useful tool for secondary quality assurance of Gamma Knife treatment plans.
机译:建立了用于伽玛刀治疗计划的独立剂量验证的形式主义。它基于近似值,单次剂量的等剂量分布在三维空间中呈椭圆形。沿三个主轴中的每一个的体模的剂量分布拟合到一个函数,该函数包含表示来自点源,聚焦外散射和平坦背景的贡献的项。针对所有四个头盔准直器,在不同的拍摄位置以及不同的头骨形状提取拟合参数。从头骨缩放仪测量中获得的患者头骨形状的33个参数是针对单个患者建模的。从31 x 31 x 31点矩阵形式的治疗体积的相对剂量是从治疗计划系统Leksell Gamma计划(LGP)中提取的。我们的模型使用与LGP中相同的输入参数来评估相对剂量,这些参数是头骨测量数据,射击位置,重量,头架的伽玛角和头盔准直器尺寸。对于29次单发病例,发现计算和LGP之间的焦点处剂量差异在-1%至2%之内。对于多发情况,计算得出的最大剂量点的值和坐标与LGP结果一致在+/- 7%和+/- 3 mm之内。通常,对于偏心位置,计算出的剂量与LGP计算值相符在+/- 10%之内。用此方法计算的50%等剂量线的尺寸和位置的结果与Leksell GammaPlan的结果非常吻合。因此,该方法可作为伽玛刀治疗计划二级质量保证的有用工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号