Consequences of air around an ionization chamber: Are existing solid phantoms suitable for reference dosimetry on an MR-linac?

机译:电离室周围空气的后果:现有的实体模型是否适合在MR直线加速器上进行参考剂量测定?

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: A protocol for reference dosimetry for the MR-linac is under development. The 1.5 T magnetic field changes the mean path length of electrons in an air-filled ionization chamber but has little effect on the electron trajectories in a surrounding phantom. It is therefore necessary to correct the response of an ionization chamber for the influence of the magnetic field. Solid phantoms are used for dosimetry measurements on the MR-linac, but air is present between the chamber wall and phantom insert. This study aimed to determine if this air influences the ion chamber measurements on the MR-linac. The absolute response of the chamber and reproducibility of dosimetry measurements were assessed on an MR-linac in solid and water phantoms. The sensitivity of the chamber response to the distribution of air around the chamber was also investigated.
机译:目的:MR-直线加速器参考剂量的协议正在开发中。 1.5 T磁场改变了充满空气的电离室中电子的平均路径长度,但对周围模型中的电子轨迹影响很小。因此,有必要针对磁场的影响校正电离室的响应。实心体模用于MR直线加速器上的剂量学测量,但是腔室壁和体模插件之间存在空气。这项研究旨在确定这种空气是否会影响MR直线加速器上离子室的测量结果。在固体和水体模型中,使用MR直线加速器评估了腔室的绝对响应和剂量测定结果的可重复性。还研究了腔室对腔室周围空气分布的响应敏感性。