首页> 外文期刊>Physics in medicine and biology. >Towards reference dosimetry for the MR-linac: Magnetic field correction of the ionization chamber reading
【24h】

Towards reference dosimetry for the MR-linac: Magnetic field correction of the ionization chamber reading

机译:迈向直线加速器参考剂量的方法:电离室读数的磁场校正

获取原文
获取原文并翻译 | 示例
           

摘要

In the UMC Utrecht a prototype MR-linac has been installed. The system consists of a 6 MV Elekta (Crawley, UK) linear accelerator and a 1.5 T Philips (Best, The Netherlands) Achieva MRI system. This paper investigates the feasibility to correct the ionization chamber reading for the magnetic field within the dosimetry calibration method described by Almond et al (1999 Med. Phys. 26 1847-70). Firstly, the feasibility of using an ionization chamber in an MR-linac was assessed by investigating possible influences of the magnetic field on NE2571 Farmer-type ionization chamber characteristics: linearity, repeatability, orientation in the magnetic field; and AAPM TG51 correction factor for voltage polarity and ion recombination. We found that these AAPM correction factors for the NE2571 chamber were not influenced by the magnetic field. Secondly, the influence of the permanent 1.5 T magnetic field on the NE2571 chamber reading was quantified. The reading is influenced by the magnetic field; therefore, a correction factor has been added. For the standardized setup used in this paper, the NE2571 chamber reading increases by 4.9% (± 0.2%) due to the transverse 1.5 T magnetic field. Dosimetry measurements in an MR-linac are feasible, if a setup-specific magnetic field correction factor (P1.5 T) for the charge reading is introduced. For the setup investigated in this paper, the P1.5 T has a value of 0.953.
机译:在UMC Utrecht中,已安装了MR-直线加速器原型。该系统由6 MV Elekta(英国克劳利)线性加速器和1.5 T飞利浦(荷兰贝斯特)的Achieva MRI系统组成。本文研究了在Almond等人(1999 Med。Phys。26 1847-70)描述的剂量学校准方法中校正磁场的电离室读数的可行性。首先,通过研究磁场对NE2571农夫型电离室特性的可能影响(线性,可重复性,磁场方向),评估了在MR直线加速器中使用电离室的可行性。 AAPM TG51校正系数,用于电压极性和离子重组。我们发现NE2571腔室的这些AAPM校正因子不受磁场的影响。其次,量化了1.5 T永久磁场对NE2571腔室读数的影响。读数受磁场影响。因此,添加了校正因子。对于本文使用的标准化设置,由于横向1.5 T磁场,NE2571腔室读数增加了4.9%(±0.2%)。如果引入用于电荷读取的特定于装置的磁场校正因子(P1.5 T),则在MR直线加速器中进行剂量测定是可行的。对于本文研究的设置,P1.5 T的值为0.953。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号