首页> 外文期刊>Medical Physics >Case report of a near medical event in stereotactic radiotherapy due to improper units of measure from a treatment planning system.
【24h】

Case report of a near medical event in stereotactic radiotherapy due to improper units of measure from a treatment planning system.

机译:由于治疗计划系统中计量单位不正确,在立体定向放射治疗中发生了一次近距离医疗事件的病例报告。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The authors hereby notify the Radiation Oncology community of a potentially lethal error due to improper implementation of linear units of measure in a treatment planning system. The authors report an incident in which a patient was nearly mistreated during a stereotactic radiotherapy procedure due to inappropriate reporting of stereotactic coordinates by the radiation therapy treatment planning system in units of centimeter rather than in millimeter. The authors suggest a method to detect such errors during treatment planning so they are caught and corrected prior to the patient positioning for treatment on the treatment machine. METHODS: Using pretreatment imaging, the authors found that stereotactic coordinates are reported with improper linear units by a treatment planning system. The authors have implemented a redundant, independent method of stereotactic coordinate calculation. RESULTS: Implementation of a double check of stereotactic coordinates via redundant, independent calculation is simple and accurate. Use of this technique will avoid any future error in stereotactic treatment coordinates due to improper linear units, transcription, or other similar errors. CONCLUSIONS: The authors recommend an independent double check of stereotactic treatment coordinates during the treatment planning process in order to avoid potential mistreatment of patients.
机译:目的:作者在此通知放射肿瘤学界,由于在治疗计划系统中线性度量单位的实施不正确而可能导致的致命错误。作者报告了这样的事件:由于放射治疗计划系统以厘米为单位而不是以毫米为单位,不恰当地报告了立体定向坐标,因此在立体定向放射治疗过程中患者几乎被误诊。作者提出了一种在治疗计划期间检测此类错误的方法,以便在将患者放置在治疗机上进行治疗之前将其捕获并纠正。方法:利用预处理成像,作者发现治疗计划系统以不正确的线性单位报告了立体定向坐标。作者已经实现了冗余,独立的立体坐标计算方法。结果:通过冗余的独立计算实现立体坐标的双重检查是简单而准确的。使用此技术将避免由于线性单位不合适,转录或其他类似错误而导致将来在立体定向治疗坐标中出现任何错误。结论:作者建议在治疗计划过程中对立体定向治疗坐标进行独立的双重检查,以避免对患者的潜在误治。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号