Scatter and crosstalk corrections for Tc-99m/I-123 dual-radionuclide imaging using a CZT SPECT system with pinhole collimators

机译:使用带有针孔准直仪的CZT SPECT系统对Tc-99m / I-123双放射性核素成像进行散射和串扰校正

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: The energy spectrum for a cadmium zinc telluride (CZT) detector has a low energy tail due to incomplete charge collection and intercrystal scattering. Due to these solid-state detector effects, scatter would be overestimated if the conventional triple-energy window (TEW) method is used for scatter and crosstalk corrections in CZT-based imaging systems. The objective of this work is to develop a scatter and crosstalk correction method for Tc-99m/I-123 dual-radionuclide imaging for a CZT-based dedicated cardiac SPECT system with pinhole collimators (GE Discovery NM 530c/570c).
机译:目的:由于不完全的电荷收集和晶体间散射,碲化镉锌(CZT)检测器的能谱具有低能量尾部。由于这些固态检测器的影响,如果在基于CZT的成像系统中使用常规的三能窗(TEW)方法进行散射和串扰校正,则散射将被高估。这项工作的目的是为带有针孔准直仪的基于CZT的专用心脏SPECT系统(GE Discovery NM 530c / 570c)开发一种Tc-99m / I-123双放射性核素成像的散射和串扰校正方法。