The dependence of the light scattering phenomena is shown in Fig. 1. In this figure, averaged response for all films from one batch lot for each channel along each direction (inplane and crossplane) is shown. As can be seen from this figure, in the direction parallel to the CCD scanner camera (crossplane direction), the response is different for each channel. For the green and blue channels, this response is not symmetric about the red channel. We would like to clarify that Perez Azorm et al? do not apply any median filter to any channel in both unexposed image and irradiated image, and the comparisons between different methods or channels are made without applying any filter.
机译:光散射现象的相关性如图1所示。在此图中,显示了一批批次中沿每个方向(面内和交叉面)的每个通道的所有薄膜的平均响应。从该图可以看出,在与CCD扫描仪相机平行的方向(横平面方向)上,每个通道的响应都不同。对于绿色和蓝色通道,此响应与红色通道不对称。我们想澄清一下Perez Azorm等人?不要在未曝光图像和辐射图像中的任何通道上应用任何中值滤镜,并且在不应用任何滤镜的情况下进行了不同方法或通道之间的比较。