首页> 外文期刊>Medical Physics >Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality.
【24h】

Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality.

机译:双源螺旋CT的时间分辨率高达3.2毫秒和75毫秒:图像重建和图像质量评估。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To present the theory for image reconstruction of a high-pitch, high-temporal-resolution spiral scan mode for dual-source CT (DSCT) and evaluate its image quality and dose. METHODS: With the use of two x-ray sources and two data acquisition systems, spiral CT exams having a nominal temporal resolution per image of up to one-quarter of the gantry rotation time can be acquired using pitch values up to 3.2. The scan field of view (SFOV) for this mode, however, is limited to the SFOV of the second detector as a maximum, depending on the pitch. Spatial and low contrast resolution, image uniformity and noise, CT number accuracy and linearity, and radiation dose were assessed using the ACR CT accreditation phantom, a 30 cm diameter cylindrical water phantom or a 32 cm diameter cylindrical PMMA CTDI phantom. Slice sensitivity profiles (SSPs) were measured for different nominal slice thicknesses, and an anthropomorphic phantom was used to assess image artifacts. Results were compared between single-source scans at pitch = 1.0 and dual-source scans at pitch = 3.2. In addition, image quality and temporal resolution of an ECG-triggered version of the DSCT high-pitch spiral scan mode were evaluated with a moving coronary artery phantom, and radiation dose was assessed in comparison with other existing cardiac scan techniques. RESULTS: No significant differences in quantitative measures of image quality were found between single-source scans at pitch = 1.0 and dual-source scans at pitch = 3.2 for spatial and low contrast resolution, CT number accuracy and linearity, SSPs, image uniformity, and noise. The pitch value (1.6 pitch 3.2) had only a minor impact on radiation dose and image noise when the effective tube current time product (mA s/pitch) was kept constant. However, while not severe, artifacts were found to be more prevalent for the dual-source pitch = 3.2 scan mode when structures varied markedly along the z axis, particularly for head scans. Images of the moving coronary artery phantom acquired with the ECG-triggered high-pitch scan mode were visually free from motion artifacts at heart rates of 60 and 70 bpm. However, image quality started to deteriorate for higher heart rates. At equivalent image quality, the ECG-triggered high-pitch scan mode demonstrated lower radiation dose than other cardiac scan techniques on the same DSCT equipment (25% and 60% dose reduction compared to ECG-triggered sequential step-and-shoot and ECG-gated spiral with x-ray pulsing). CONCLUSIONS: A high-pitch (up to pitch = 3.2), high-temporal-resolution (up to 75 ms) dual-source CT scan mode produced equivalent image quality relative to single-source scans using a more typical pitch value (pitch = 1.0). The resultant reduction in the overall acquisition time may offer clinical advantage for cardiovascular, trauma, and pediatric CT applications. In addition, ECG-triggered high-pitch scanning may be useful as an alternative to ECG-triggered sequential scanning for patients with low to moderate heart rates up to 70 bpm, with the potential to scan the heart within one heart beat at reduced radiation dose.
机译:目的:介绍用于双源CT(DSCT)的高螺距,高时间分辨率螺旋扫描模式的图像重建理论,并评估其图像质量和剂量。方法:通过使用两个X射线源和两个数据采集系统,可以使用不超过3.2的螺距值来获取每张图像的名义时间分辨率高达龙门架旋转时间四分之一的螺旋CT检查。但是,取决于间距,此模式的扫描视场(SFOV)最大限于第二个检测器的SFOV。使用ACR CT认证体模,直径30 cm的圆柱水体模或32 cm直径的PMMA CTDI体模评估了空间和低对比度分辨率,图像均匀性和噪声,CT数的准确性和线性以及辐射剂量。测量了不同标称切片厚度的切片灵敏度曲线(SSP),并使用了拟人模型来评估图像伪影。比较了间距为1.0的单源扫描和间距为3.2的双源扫描的结果。此外,使用移动的冠状动脉体模评估了ECCT触发的DSCT高音螺旋扫描模式的图像质量和时间分辨率,并与其他现有的心脏扫描技术进行了比较,评估了辐射剂量。结果:在间距= 1.0的单源扫描和间距= 3.2的双源扫描之间,在空间和低对比度分辨率,CT数精度和线性,SSP,图像均匀性和噪声。当有效管电流时间乘积(mA s / pitch)保持恒定时,螺距值(1.6螺距3.2)仅对辐射剂量和图像噪声有很小的影响。但是,尽管不严重,但当结构沿z轴显着变化时,尤其是对于头部扫描,发现在双源间距= 3.2扫描模式下,伪影更为普遍。使用ECG触发的高音高扫描模式获取的运动冠状动脉体模的图像在视觉上没有以60 bpm和70 bpm的心率运动伪影。但是,随着心跳加快,图像质量开始下降。在相同的图像质量下,在相同的DSCT设备上,ECG触发的高音高扫描模式显示出比其他心脏扫描技术更低的辐射剂量(与ECG触发的连续步进拍摄和ECG-相比,剂量减少了25%和60%带有X射线脉冲的门控螺旋)。结论:高音调(最高音调= 3.2),高时间分辨率(最高75 ms)双源CT扫描模式产生的图像质量与使用更典型音调值的单源扫描相比具有相同的图像质量(音高= 1.0)。总体采集时间的减少可为心血管,创伤和儿科CT应用提供临床优势。此外,对于心率高达70 bpm的中低心率患者,ECG触发的高音高扫描可能会替代ECG触发的顺序扫描,并有可能以降低的辐射剂量在一次心跳内扫描心脏。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号