首页> 外文期刊>Magnetic resonance in medical sciences: MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine >A pitfall of the volume rendering method with 3D time-of-flight MRA: A case of a branching vessel at the aneurysm neck
【24h】

A pitfall of the volume rendering method with 3D time-of-flight MRA: A case of a branching vessel at the aneurysm neck

机译:3D飞行时间MRA的体积渲染方法的一个陷阱:动脉瘤颈处分支血管的情况

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We present a case in which the origin of the branching vessel at the aneurysm neck was observed at the wrong place on the volume rendering method (VR) with 3D time-of-flight MRA (3D-TOF-MRA) with 3-Tesla MR system. In 3D-TOF-MRA, it is often difflcult to observe the origin of the branching vessel, but it is unusual for it to be observed in the wrong place. In the planning of interventional treatment and surgical procedures, false recognition, as in the unique case in the present report, is a serious problem. Decisions based only on VR with 3D-TOF-MRA can be a cause of suboptimal selection in clinical treatment.
机译:我们提出了一种情况,在使用3D飞行时间MRA(3D-TOF-MRA)和3-Tesla MR的体积渲染方法(VR)上,在错误的位置观察到了动脉瘤颈部分支血管的起源系统。在3D-TOF-MRA中,通常很难观察到分支血管的起源,但是在错误的位置观察到它是不寻常的。在介入治疗和外科手术的计划中,与本报告中的独特情况一样,错误识别是一个严重的问题。仅基于VR和3D-TOF-MRA的决策可能是临床治疗选择不佳的原因。