首页> 外文期刊>Canadian journal of anesthesia: Journal canadien d'anesthesie >An assessment of the luminance and light field characteristics of used direct laryngoscopes.
【24h】

An assessment of the luminance and light field characteristics of used direct laryngoscopes.

机译:对使用过的直接喉镜的亮度和光场特性的评估。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: To determine luminance and light field characteristics and the effect of residual battery potential and luminance on light colour temperature in our used Macintosh #3 and #4 fibre-light (FLB) and bulb-light (BLB) laryngoscopes. METHODS: We used a power supply to provide laryngoscopes with potentials equivalent to those most commonly measured in the handles in use in our OR. Measurements were made under controlled, constant, conditions using a Pentax digital spotmeter (luminance) or a Minolta Color III colour temperature meter (light colour). Colour measurements were made while increasing the power source potential from 2-3 volts (v) in increments of 0.1 v. Light field measurements were made with a mm increment ruler mounted on the base of the test fixture. RESULTS: At 2.5 and 2.8 v respectively, the #3 FLB produced luminance values of 23.9 +/- 11.4 and 41.7 +/- 17.2 cd x m(-2) (mean +/- SD), and the #4 FLB produced 58.6 +/- 21.4 and 90.9 +/- 32.2 cd x m(-2). Increasing potential increased luminance values (P < 0.001) and the #4FLB produced higher luminance values (P < 0.001). BLB produced higher luminance values than did FLB across all comparisons (P < 0.001). As potentials and luminance values decreased, light temperature was reduced (P < 0.001). There were no differences in light field dimensions noted in any comparison. CONCLUSION: Fifteen percent of the BLB did not meet the minimum luminance for laryngoscopy of 100 cd x m(-2), 92% of the FLB did not meet that same standard.
机译:目的:在我们使用的Macintosh#3和#4光纤(FLB)和灯泡(BLB)喉镜中,确定亮度和光场特性以及剩余电池电势和亮度对光色温的影响。方法:我们使用电源为喉镜提供的电位与手术室使用的手柄中最常用的电位相等。使用Pentax数字斑点计(亮度)或Minolta Color III色温计(浅色)在受控的恒定条件下进行测量。进行颜色测量时,将电源电势从2-3伏(v)以0.1 v的增量增加。使用安装在测试夹具底座上的mm增量尺测量光场。结果:分别在2.5和2.8 v时,#3 FLB产生的亮度值为23.9 +/- 11.4和41.7 +/- 17.2 cd xm(-2)(平均+/- SD),而#4 FLB产生的亮度值为58.6 + 21.4和90.9 +/- 32.2 cd xm(-2)。潜在的增加的亮度值(P <0.001)和#4FLB产生更高的亮度值(P <0.001)。在所有比较中,BLB均比FLB产生更高的亮度值(P <0.001)。随着电势和亮度值的降低,光温降低(P <0.001)。在任何比较中,光场尺寸都没有差异。结论:15%的BLB不符合100 cd x m(-2)的喉镜最低照度,92%的FLB不符合相同的标准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号