首页> 外文期刊>Flow Measurement and Instrumentation >A practical method to assess the performance of gravimetric flow test rigs by using the timing error
【24h】

A practical method to assess the performance of gravimetric flow test rigs by using the timing error

机译:利用定时误差评估重力流量试验台性能的实用方法

获取原文
获取原文并翻译 | 示例
           

摘要

This paper proposes an improved method to assess the performance of a gravimetric flow test rig. It addresses the study of two of the most important measurement systems: the weighing scale and all mass-related corrections, and the time measurement system. By performing a regular diverter test according to ISO 4185:1980 several times and under different conditions, it was possible to determine not only the statistical variances of the diverter system, but also the variances of the meter under test and of the flow test facility separately. The proposed method allows also under special conditions the determination of systematic errors in the mass determination. The method has been successfully applied at the large water flow calibration facilities of PTB in Braunschweig and Berlin and is recommended as a routine to test the performance of gravimetric flow test facilities. (C) 2014 The Authors. Published by Elsevier Ltd.
机译:本文提出了一种改进的方法来评估重量流量试验台的性能。它解决了对两个最重要的测量系统的研究:秤和所有与质量相关的校正,以及时间测量系统。通过根据ISO 4185:1980在不同条件下多次执行常规分流器测试,不仅可以确定分流器系统的统计差异,还可以分别确定被测仪表和流量测试设施的差异。所提出的方法还允许在特殊条件下确定质量确定中的系统误差。该方法已成功地在不伦瑞克和柏林的PTB大型水流量校准设施中应用,推荐作为常规方法来测试重量流量测试设施的性能。 (C)2014作者。由Elsevier Ltd.发布