首页> 外文期刊>Fluid Phase Equilibria >Model-free calibration of Raman measurements of reactive systems: Application to monoethanolamine/water/CO2
【24h】

Model-free calibration of Raman measurements of reactive systems: Application to monoethanolamine/water/CO2

机译:反应系统拉曼测量的无模型校准:在单乙醇胺/水/二氧化碳中的应用

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We present a method to calibrate Raman measurements of reactive systems without relying on thermodynamic models. Raman spectroscopy is able to detect multiple species even in reactive mixtures. Classical calibration methods require the knowledge of concentrations of all species as input. Typically, concentrations of intermediates cannot be prepared and fixed independently and are therefore calculated using thermodynamic models. The quality of the results thus depends on the accuracy of the thermodynamic model. In this paper, we present a method based only on stoichiometric balances and electroneutrality. By avoiding the use of thermodynamic models in the calibration step, the risk of overfitting spectroscopic data to a thermodynamic model is avoided. The presented method is demonstrated for the reactive system monoethanolamine, water, and CO2 and is validated by a thermodynamic model taken from the literature. While the model-free calibration is demonstrated for Raman spectroscopy in this work, the approach is generic should thus be applicable to most spectroscopic techniques. (C) 2015 Elsevier B.V. All rights reserved.
机译:我们提出了一种无需依赖热力学模型即可校准反应堆系统拉曼测量的方法。拉曼光谱仪甚至可以在反应性混合物中检测多种物质。经典的校准方法需要了解所有物种的浓度作为输入。通常,中间体的浓度无法独立制备和固定,因此需要使用热力学模型进行计算。因此,结果的质量取决于热力学模型的准确性。在本文中,我们提出了一种仅基于化学计量平衡和电子中性的方法。通过避免在校准步骤中使用热力学模型,避免了将光谱数据过度拟合为热力学模型的风险。该方法针对单乙醇胺,水和二氧化碳的反应体系进行了论证,并通过取自文献的热力学模型进行了验证。尽管在这项工作中对拉曼光谱进行了无模型校准,但该方法是通用的,因此应适用于大多数光谱技术。 (C)2015 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号