首页> 外文期刊>Displays: Technology and Applications >Analyses of system error in the measurement of liquid crystal empty cell gap by means of interferometry
【24h】

Analyses of system error in the measurement of liquid crystal empty cell gap by means of interferometry

机译:基于干涉测量法测量液晶空池间隙的系统误差分析

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Influences on system error and relative standard deviation in the measurement of liquid crystal empty cell gap are analyzed in great detail. Great differences exist in system error and standard deviation among measuring methods. Momentum Scanning Interferometry (MST) using the P wave can keep system error at a small level. ITO and PI layers are the main sources of system error, but their influence on MSI using the P wave is the least among different methods. Based on these analyses, an effective solution, MSI using sloping tiny beam, is presented. It can be used to measure the distribution of cell gap on line and is of high resolution, good repeatability, good stability and high precision. (C) 2000 Elsevier Science B.V. All rights reserved. References: 2
机译:详细分析了液晶空池间隙测量对系统误差和相对标准偏差的影响。不同测量方法在系统误差和标准偏差方面存在很大差异。使用P波的动量扫描干涉测量(MST)可以将系统误差保持在较小的水平。ITO 和 PI 层是系统误差的主要来源,但在不同方法中,它们对使用 P 波的 MSI 的影响最小。基于这些分析,提出了一种有效的解决方案,即使用倾斜微小光束的MSI。可用于在线测量细胞间隙分布,分辨率高,重复性好,稳定性好,精度高。(C) 2000 Elsevier Science B.V.保留所有权利。[参考资料: 2]

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号