首页> 外文期刊>Gait & posture >Improving repeatability of setting volume origin and coordinate system for 3D gait analysis
【24h】

Improving repeatability of setting volume origin and coordinate system for 3D gait analysis

机译:提高设置体积原点和坐标系以进行3D步态分析的可重复性

获取原文
获取原文并翻译 | 示例
           

摘要

This study investigated the variability arising from replacement of different calibration objects. A new calibration object (L-Frame) was instigated which consisted of four markers screwed into fixed positions in the floor, to ensure consistency between replacements. The variability of a healthy subject's joint moments from replacement of different calibration objects (5 marker wand and L-Frame) was compared to the subject's stride-to-stride repeatability (SSR). Overall the L-Frame calibration object demonstrated the least amount of variation during object replacement for both sagittal and coronal joint moments. In particular for the coronal ankle moment the 5 marker wand positioned at the end of the force-plate alignment had a variability of 81% of SSR, while the L-Frame showed only 3% of SSR. Utilising the L-Frame to set the volume origin and global coordinate system achieved the most consistent results, which is important for assuring reliable measurements during 3-Dimensional Gait Analysis (3DGA).
机译:这项研究调查了因更换不同校准对象而引起的变异性。提出了一个新的校准物体(L形框架),该物体由四个标记拧入地板上的固定位置组成,以确保更换之间的一致性。将健康受试者的关节力矩因更换不同的校准物体(5个标记棒和L型框架)而发生的变化与受试者的跨步重复性(SSR)进行了比较。总体而言,L形架标定物体在矢状和冠状关节力矩的物体更换过程中表现出的变化最小。特别是对于冠状踝力矩,位于力板对准末端的5个标记棒的SSR变异度为81%,而L形支架的SSR变异度仅为3%。利用L框架设置体积原点和全局坐标系可获得最一致的结果,这对于确保3维步态分析(3DGA)期间的可靠测量非常重要。