Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes

机译:填充硅油的眼睛中的IOL主镜和A扫描浸入式生物测定法的准确性和可靠性

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PurposeTo compare the accuracy and reliability of intraocular lens (IOL) master and A-scan immersion biometry in silicone oil (SO)-filled eyes.MethodsA prospective, consecutive, nonrandomized study was performed in 34 SO-filled eyes of 34 patients, who underwent a pars plana vitrectomy, with SO removal and cataract surgery, as well as IOL implantation. Both IOL master and immersion A-scan were performed to measure the axial length (AXL) before SO removal. Three months after removal of the SO, AXL measurements using IOL master and refraction was performed. Accuracy of the two techniques was determined by a mean postoperative AXL using an IOL master and reliability was determined by mean actual postoperative refractive error.ResultsPreoperative mean AXL was 23.91±0.24 mm (range 21.33-28.61 mm) and 23.71±0.59 mm (range 19.27-36.18 mm) by IOL master and A-scan immersion, respectively. Postoperative mean AXL by IOL master was 23.90±0.23 mm (range 21.58-27.94 mm), which showed a statistically significant difference from the preoperative mean AXL by A-scan immersion (P0.005). The AXL measurement by IOL master also was more accurate than A-scan immersion by Pearsons correlation (0.966 vs 0.410). For reliability of the two techniques, the predictive postoperative refractive error in A-scan immersion (mean 1.79±1.04 D, range 14.62 to 16.41 D) was greater than that in IOL master (mean 0.60±0.23 D, range 2.74 to 2.33 D), with a statistically significant difference (P0.049).ConclusionIOL master had more accuracy and less deviation in predictive postoperative refractive error than A-scan immersion in SO-filled eyes.
机译:目的比较经硅油(SO)填充的眼睛的人工晶状体(IOL)主镜和A扫描浸入式生物测定法的准确性和可靠性。方法对34例接受SO的人的34眼进行了前瞻性,连续,非随机研究进行pars平面玻璃体切除术,并进行SO去除和白内障手术以及IOL植入。在去除SO之前,均进行了IOL主扫描和浸没A扫描以测量轴向长度(AXL)。去除SO后三个月,使用IOL主镜和折射仪进行AXL测量。两种技术的准确度由术后使用IOL母板的平均AXL来确定,而可靠性则由术后平均实际屈光误差来确定。 -36.18毫米)分别通过IOL主镜和A扫描浸入。 IOL大师的术后平均AXL为23.90±0.23 mm(范围21.58-27.94 mm),与A扫描浸入术前的平均AXL相比具有统计学显着性差异(P0.005)。通过Pearsons相关性,IOL主设备进行的AXL测量也比A扫描浸入更准确(0.966对0.410)。为了这两种技术的可靠性,A扫描​​浸入的预测术后屈光不正(平均1.79±1.04 D,范围14.62至16.41 D)要大于IOL母版(平均0.60±0.23 D,范围2.74至2.33 D)。 ,具有统计学显着性差异(P0.049)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号