Letter to the editor re: low-dose computed tomography of the paranasal sinus and facial skull using a high-pitch dual-source system-first clinical results.

机译:致编辑的信重新:使用高音高双源系统对鼻旁窦和面部颅骨进行低剂量计算机断层扫描的第一个临床结果。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Dear Editor, We read with great interest the recent work by B Schell et al. [1], who used a dual-source Flash CT to examine the paranasal sinus in patients, comparing a high pitch and low pitch protocol in terms of dose and image quality. The authors of this study concluded that diagnostic image quality can be maintained while cutting the radiation exposure down to approximately half (Volume CT dose index: 7.22 mGy vs 2.77 mGy) using the dual-source high pitch mode [1]. However, we believe this conclusion may be misleading for the readers. Changing the pitch should not significantly affect the dose on this system as Siemens (Forchheim, Germany) uses a Sure View technique (effective mAs rather than mAs) to maintain image quality and dose regardless of selected pitch.
机译:亲爱的编辑,我们非常感兴趣地阅读了B Schell等人的最新著作。 [1],他使用双源Flash CT检查患者的鼻旁窦,在剂量和图像质量方面比较了高音高和低音高的方案。这项研究的作者得出的结论是,使用双源高螺距模式可以将放射线的照射量减少大约一半(CT剂量指数:7.22 mGy对2.77 mGy),从而可以保持诊断图像的质量[1]。但是,我们认为这个结论可能会误导读者。更改间距不会对本系统的剂量产生重大影响,因为西门子(德国福希海姆)使用Sure View技术(有效mAs而非mAs)来保持图像质量和剂量,而不管选择的间距如何。