首页> 外文期刊>European radiology >Single-source dual-energy CT angiography with reduced iodine load in patients referred for aortoiliofemoral evaluation before transcatheter aortic valve implantation: impact on image quality and radiation dose.
【24h】

Single-source dual-energy CT angiography with reduced iodine load in patients referred for aortoiliofemoral evaluation before transcatheter aortic valve implantation: impact on image quality and radiation dose.

机译:经导管主动脉瓣植入术前转入主动脉股动脉评估患者的碘负荷降低的单源双能CT血管造影:对图像质量和放射剂量的影响。

获取原文
获取原文并翻译 | 示例

摘要

? Dual-energy AICTA before TAVI results in a 50 % reduction of iodine load. ?The reduction of iodine load maintains sufficient image quality despite increased noise. ? Using 375 mA in dual-energy mode results in a reduction of radiation dose. ? A high tube current setting (600 mA) should be used in overweight patients.
机译:? TAVI之前的双能AICTA可将碘负荷降低50%。 ?尽管增加了噪音,但减少碘负荷仍可保持足够的图像质量。 ?在双能模式下使用375 mA电流可以减少辐射剂量。 ?超重患者应使用较高的电子管电流设置(600 mA)。