Ordinary surface ECG electrodes accurately reflect cardiac electric activity at hypothermia.

机译:普通的表面ECG电极可准确反映体温过低时的心脏电活动。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: It has been claimed that needle electrodes can be a useful means to detect weak ECG signals in cases of accidental hypothermia. METHODS: Four pigs were cooled by immersion in ice water, followed by direct cooling of the blood through an extracorporeal circulation system until the core temperature was lowered to 12 degrees C and surface-measured ECG indicated asystole. Following cooling, the pigs were rewarmed and weaned from extracorporeal circulation if possible. ECG and interelectrode impedance were measured between surface electrodes, needle electrodes and electrodes sewn to the epicardium during the cooling and rewarming procedure. RESULTS: Needle and surface electrodes showed exactly the same ECG whatever the temperature of the skin or the core was. The impedance varied only slightly with temperature and could not explain the disappearance of surface ECG. The QRS wave amplitude showed the greatest sensitivity to temperature, disappearing completely before the P-wave disappeared. The P-wave showed the least sensitivity, and was the last wave to disappear, indicating that the sinus node is the most resistant part of the heart to cooling. Between 19 and 17 degrees C, a commercial monitor indicated asystole although P-waves could be seen in the ECG and atrial contractions could be visually observed on the heart. CONCLUSION: Surface electrodes had a similar high accuracy to indicate electric activity as needle electrodes. Higher amplification and reduction of the timebase made it possible to detect ECG in a situation where asystole was indicated by commercial monitors.
机译:背景:有人声称,在意外体温过低的情况下,针电极可能是检测弱心电信号的有用手段。方法:将四头猪浸入冰水中冷却,然后通过体外循环系统直接冷却血液,直到核心温度降至12摄氏度,并且经表面测量的心电图显示心脏停搏。冷却后,如果可能的话,将猪重新加热并使其脱离体外循环。在冷却和加热过程中,测量了表面电极,针状电极和缝在心外膜上的电极之间的ECG和电极间阻抗。结果:无论皮肤或核心温度如何,针电极和表面电极均显示完全相同的ECG。阻抗仅随温度变化很小,不能解释表面心电图的消失。 QRS波振幅显示出对温度的最大敏感性,在P波消失之前完全消失。 P波显示的敏感性最低,并且是最后消失的波,表明窦房结是心脏对冷却的抵抗力最大的部分。在19到17摄氏度之间,尽管在ECG中可以看到P波并且可以在心脏上肉眼观察到心房收缩,但是商用监护仪仍显示出了心搏停止。结论:表面电极具有与针状电极类似的高精度,可指示电活动。更高的放大倍数和时基的减少使得在商业监护仪指示心搏停止的情况下检测ECG成为可能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号