Three-dimensional finite element analysis was performed for thin hydroxyapatite (HA) coated and titanium dental implants to study the effects on stress/strain distribution in the mandible with application of axial and oblique loads. The implants were of screw and cylinder types. With an axial load, the maximum equivalent bone stresses in the titanium implants were 21.5 and 29.0 MPa for the cylinder and screw types respectively, and the stress and strain distributions differed. For the cylinder type, the highest stress was located at the implant base, and for the screw type, it was located at the top edge of the first thread within the cortical bone. For the HA-coated cylinder and screw implants, the maximum equivalent bone stresses were 7.1 and 7.2 MPa respectively. The stress and strain distributions were similar, and the highest stress was located on the upper side of the cortical bone around the implant neck for both implants. Of the implants examined, the screw type HA-coated implant had the most uniform stress distribution in bone.
机译:对羟基磷灰石(HA)涂层和钛牙种植体进行了三维有限元分析,以研究轴向和斜向载荷对下颌骨中应力/应变分布的影响。植入物为螺钉和圆柱体类型。在轴向载荷下,圆柱型和螺钉型钛植入物的最大等效骨应力分别为21.5 MPa和29.0 MPa,应力和应变分布不同。对于圆柱型,最高应力位于植入物的底部,对于螺钉型,最高应力位于皮质骨内第一根螺纹的顶部边缘。对于HA涂层的圆柱体和螺钉植入物,最大等效骨应力分别为7.1 MPa和7.2 MPa。应力和应变分布相似,两种植入物的最高应力都位于植入物颈部周围的皮质骨上侧。在所检查的植入物中,螺钉型HA涂层植入物在骨骼中的应力分布最均匀。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号