首页> 外文期刊>Europace: European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology >Morphology discrimination in implantable cardioverter-defibrillators: consistency of template match percentage during atrial tachyarrhythmias at different heart rates.
【24h】

Morphology discrimination in implantable cardioverter-defibrillators: consistency of template match percentage during atrial tachyarrhythmias at different heart rates.

机译:植入式心脏复律除颤器的形态学辨别:在不同心率的房性快速性心律失常中模板匹配百分比的一致性。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Morphology discrimination (MD) in implantable cardioverter-defibrillators (ICDs) is based on the comparison of the ventricular electrogram during tachycardia with a stored reference template obtained during baseline rhythm. However, the effect of heart rate on the template match percentage during supraventricular tachyarrhythmias (SVT) is not known. The purpose of this study was to evaluate the performance of the template match percentage during SVT at different heart rates. METHODS AND RESULTS: Stored electrograms of 868 tachyarrhythmias from 88 patients with a dual-chamber ICD (St Jude Medical, USA) were analysed by the investigators. The effect of heart rate on template match percentage was estimated by regression analysis. For performance measures, data were corrected for multiple episodes in a patient by using the generalized estimating equation method. The mean template match percentage was 86.6 +/- 22.2% (median 100%) for SVT episodes. No significant differences in template match percentage between fast [ventricular cycle length (CL) 300-350 ms] and slow (ventricular CL >400 ms) SVTs were observed (85.4 +/- 27.0 vs. 87.1 +/- 19.7%). Using nominal settings, MD alone provided sensitivity and specificity of 70.2% and 89.4% overall, respectively. Morphology discrimination in conjunction with rate branch analysis, sudden onset, and stability yielded sensitivity and specificity of 98.5% and 91.2%, respectively. CONCLUSION: Morphology discrimination has a consistently high template match percentage during SVTs, which is independent of ventricular CL. The consistent high match percentage results in high specificity for arrhythmia discrimination.
机译:背景:植入式心脏复律除颤器(ICD)中的形态学辨别(MD)是基于心动过速期间心室电描记图与基线节律期间获得的参考模板的比较。但是,心律对室上性快速性心律失常(SVT)期间模板匹配百分比的影响尚不清楚。这项研究的目的是评估在不同心率的SVT期间模板匹配百分比的性能。方法和结果:研究者分析了88例双腔ICD患者(美国圣裘德医疗公司)的868例快速性心律失常的电图。心率对模板匹配率的影响通过回归分析估算。对于绩效指标,使用广义估计方程法对患者多次发作的数据进行了校正。 SVT发作的平均模板匹配百分比为86.6 +/- 22.2%(中位数100%)。在快速[心室周期长度(CL)300-350 ms]和慢速(心室CL> 400 ms)SVT之间,模板匹配百分比没有显着差异(85.4 +/- 27.0 vs. 87.1 +/- 19.7%)。使用标称设置,仅MD单独提供的总体敏感性和特异性分别为70.2%和89.4%。形态学鉴别与速率分支分析,突然发作和稳定性相结合,分别产生了98.5%和91.2%的敏感性和特异性。结论:形态学鉴别在SVT期间具有始终较高的模板匹配率,这与心室CL无关。一致的高匹配百分比会导致对心律失常的高度特异性。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号