首页> 外文期刊>Brachytherapy >Sensitivity of clinically relevant dosimetric parameters to contouring uncertainty in postimplant dosimetry of low-dose-rate prostate permanent seed brachytherapy
【24h】

Sensitivity of clinically relevant dosimetric parameters to contouring uncertainty in postimplant dosimetry of low-dose-rate prostate permanent seed brachytherapy

机译:低剂量率前列腺永久性种子近距离放疗的植入后剂量学中临床相关剂量学参数对轮廓不确定性的敏感性

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: There is strong evidence relating postimplant dosimetry for low-dose-rate prostate seed brachytherapy to local control rates. The delineation of the prostate on CT images, however, represents a challenge due to the lack of soft-tissue contrast to identify the prostate borders. This study aims at quantifying the sensitivity of prostate V-100 and D-90 to contouring uncertainty as clinically relevant parameters for evaluation of target coverage in postimplant dosimetry.
机译:目的:有强有力的证据证明低剂量率前列腺种子近距离放射治疗的植入后剂量与局部控制率有关。然而,由于缺乏用于识别前列腺边界的软组织造影剂,因此在CT图像上描绘前列腺代表了一项挑战。这项研究旨在量化前列腺V-100和D-90对轮廓不确定性的敏感性,作为临床相关参数,用于评估植入后剂量测定中的靶标覆盖率。