Film-screen magnification versus electronic magnification and enhancement of digitized contact mammograms in the assessment of subtle microcalcifications.

机译:胶片屏幕放大倍率与电子放大倍数的比较,以及数字化乳房X线照片的增强,可用于评估微小的微钙化。

获取原文
获取原文并翻译 | 示例

摘要

RATIONALE AND OBJECTIVES: To compare information drawn from magnification mammography with that extracted from electronic magnification, processing, and display of the digitized contact images. METHODS: Contact and magnification images of a mammographic statistical phantom were obtained. The magnification films versus the computer-enhanced, digitized images of the corresponding contact mammograms were separately presented to three observers. Receiver operating characteristic analysis was used to compare lesion detectability. The contact and magnification mammograms of 86 patients with subtle microcalcifications were also studied. The breast imaging reporting and data system (BI-RADS) scheme was used to compare the magnification patient films versus the corresponding digitized contact images. Differences in mammographic assessment were evaluated by using the kappa statistic. The dose to breast tissue from contact and magnification mammography was measured to evaluate dose reduction in instances where magnification mammography was to be avoided. RESULTS: Lesion detectability was found to be similar when either the digitized film image or the magnification hard-copy film was inspected. Interpretation of patient images by inspection of the contact and magnification screen-film mammograms on a view-box was in excellent agreement with that yielded by inspection of the contact image on a view-box and the computer-enhanced, digitized contact image on a display monitor. CONCLUSIONS: Electronic magnification and processing of the digitized contact image may provide valuable information concerning subtle microcalcifications, rendering magnification mammography unnecessary for many patients with such lesions.
机译:理由和目的:将放大的乳腺X线摄影术中获得的信息与从数字化的接触图像的电子放大率,处理和显示中提取的信息进行比较。方法:获得乳房X光照相术统计体模的接触和放大图像。放大胶片与相应的接触乳房X线照片的计算机增强的数字化图像分别呈现给三个观察者。接受者操作特征分析用于比较病变的可检测性。还对86例微钙化患者的接触和放大X线照片进行了研究。乳房成像报告和数据系统(BI-RADS)方案用于比较放大的患者胶片与相应的数字化接触图像。乳腺X线摄影评估的差异通过kappa统计进行评估。测量了接触和放大乳腺摄影对乳房组织的剂量,以评估在应避免使用放大乳腺摄影的情况下的剂量减少情况。结果:当检查数字化胶片图像或放大硬拷贝胶片时,发现病变的可检测性相似。通过查看观察盒上的接触和放大屏幕胶片乳房X线照片来解释患者图像,与通过查看观察盒上的接触图像和显示器上的计算机增强的数字化接触图像得到的解释非常一致。监控。结论:数字化接触图像的电子放大和处理可能会提供有关微细钙化的有价值的信息,这使得许多具有此类病变的患者无需进行乳房X线摄影。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号