首页> 外文期刊>International Orthopaedics >Increased loosening of cemented straight stem prostheses made from titanium alloys. An analysis and comparison with prostheses made of cobalt-chromium-nickel alloy.
【24h】

Increased loosening of cemented straight stem prostheses made from titanium alloys. An analysis and comparison with prostheses made of cobalt-chromium-nickel alloy.

机译:钛合金制成的硬质合金直柄假体的松动增加。与钴铬镍合金制成的假体的分析和比较。

获取原文
获取原文并翻译 | 示例
       

摘要

We studied the rate of aseptic loosening of three different types of femoral stems in primary total hip replacement. After a median follow-up of 10.2 years 4/147 CoCrNi (SS77) straight stems (type M.E. Muller) were revised. After a median follow-up of 7.7 years 32/239 Ti-6A1-7Nb (SS77) were revised and after a median follow-up of 5.2 years 52/203 SLS Titanium alloy stems were revised. Whereas the first two stems are of identical design (smooth-blasted, anterior and posterior collar), the SLS stem design is different. Surgical procedure and cementing technique have remained unchanged. There is a significantly higher risk of failure for smaller titanium stem sizes and in males and patients who are physically active. This indicates that the greater elasticity of the Titanium alloy is one of the factors responsible for loosening.
机译:我们研究了初次全髋置换中三种不同类型股骨柄的无菌性松动率。中位随访10.2年后,对4/147 CoCrNi(SS77)直茎(M.E. Muller型)进行了修订。中位随访7.7年后,对32/239 Ti-6A1-7Nb(SS77)进行了修订,中位随访了5.2年后,对52/203 SLS钛合金杆进行了修订。前两个茎具有相同的设计(喷砂,前和后领),但SLS茎的设计却不同。手术程序和固井技术保持不变。钛杆尺寸较小以及男性和体育锻炼患者的失败风险明显更高。这表明钛合金的更大的弹性是造成松弛的因素之一。