摘要

Fluke Calibration (formerly Hart Scientific) in American Fork, Utah, USA is a manufacturer of temperature calibration instruments. The company manufactured reference standard gold versus platinum (Au-Pt) thermocouples from 1992 to about 2002. Manufacturing was halted in 2002 because a trend of poor curve-fit results was observed in new batches of wire. After reviewing the possible sources of the problem, it was decided to sample wire from multiple manufacturers and investigate ways to make the curve-fit work better. This paper presents the results from the study of the wire and a characterization technique to help improve characterization of thermocouples made with lower purity wire. Calibration results from NIST SRM material and older Fluke thermocouples are included as well to provide a means of comparison of contemporary wire to NIST SRM era wire.
机译:位于美国犹他州美国叉的Fluke Calibration(以前称为Hart Scientific)是温度校准仪器的制造商。该公司从1992年到2002年左右制造了参考标准的金对铂(Au-Pt)热电偶。由于在新一批焊丝中发现了曲线拟合结果较差的趋势,因此2002年停止了生产。在审查了可能的问题根源之后,决定从多个制造商处取样导线,并研究使曲线拟合更好地工作的方法。本文介绍了电线的研究结果和一种表征技术,可帮助改善由纯度较低的电线制成的热电偶的表征。还包括NIST SRM材料和较旧的Fluke热电偶的校准结果,以提供将现代电线与NIST SRM时代电线进行比较的方法。