A modified auxiliary excitation signal for achieving unidirectional ion ejection in quadrupole ion trap mass spectrometers operating in the resonance ejection mode

机译:修改后的辅助激励信号,用于在以共振喷射模式运行的四极杆离子阱质谱仪中实现单向离子喷射

获取原文
获取原文并翻译 | 示例
       

摘要

This paper proposes a technique to cause unidirectional ion ejection in a quadrupole ion trap mass spectrometer operated in the resonance ejection mode. In this technique a modified auxiliary dipolar excitation signal is applied to the endcap electrodes. This modified signal is a linear combination of two signals. The first signal is the nominal dipolar excitation signal which is applied across the endcap electrodes and the second signal is the second harmonic of the first signal, the amplitude of the second harmonic being larger than that of the fundamental.
机译:本文提出了一种在共振喷射模式下运行的四极杆离子阱质谱仪中引起单向离子喷射的技术。在该技术中,将修改的辅助偶极激励信号施加到端盖电极。该修改后的信号是两个信号的线性组合。第一信号是施加在端帽电极上的标称偶极激励信号,第二信号是第一信号的第二谐波,第二谐波的幅度大于基频。