Comparison of retinal nerve fibre layer thickness measurements calculated by the optic nerve head map (NHM4) and RNFL3.45 modes of spectral-domain optical coherence tomography (RTVue-100).

机译:通过视神经头图(NHM4)和光谱域光学相干断层扫描(RTVue-100)的RNFL3.45模式计算的视网膜神经纤维层厚度测量结果的比较。

获取原文
获取原文并翻译 | 示例
       

摘要

AIM: To evaluate and compare retinal nerve fibre layer (RNFL) thickness measured by direct scanning (RNFL3.45 mode) and re-sampling from datasets (NHM4 mode) of RTVue-100 optical coherence tomography (OCT). METHODS: Thirty-six healthy subjects and 76 subjects with glaucoma were imaged with Stratus OCT (fast RNFL mode, RNFL3) and RTVue-100 OCT (NHM4 (RNFL1) and RNFL3.45 (RNFL2) modes). Measurement reproducibility was assessed in NHM4 and RNFL3.45 modes of RTVue-100 OCT (intraclass correlation coefficient (ICC)). Agreement between different RNFL measurements was analysed by Bland-Altman plot. The areas under the receiver operating characteristic (ROC) curves (AUCs) for discrimination between healthy and glaucoma were compared between the different RNFL measurements. RESULTS: Both NHM4 and RNFL3.45 modes showed excellent measurement reproducibilities (ICC 0.831-0.978). RNFL thicknesses by two different modes of the RTVue-100 OCT, and by the Stratus OCT, were correlated in all sectors, including average. There was no significant difference between RNFL1 and RNFL2 data. However, RNFL thickness in glaucomatous eyes by the RTVue-100 OCT was significantly greater than that measured by Stratus OCT. RNFL thicknesses determined by the two different modes of RTVue-100 OCT were in excellent agreement (95% limits of agreement -6.53 to 6.95 mum). All three RNFL measurements showed good glaucoma discrimination ability (AUC = RNFL1 0.970, RNFL2 0.962, RNFL3 0.971). CONCLUSIONS: RNFL thickness determined by direct scanning and re-sampling from datasets of RTVue-100 OCT were in good agreement. However, both measurements in glaucomatous subjects were significantly different from those with the Stratus OCT. Those findings should be considered when a patient is followed-up using different types of OCT.
机译:目的:评估和比较通过直接扫描(RNFL3.45模式)和从RTVue-100光学相干断层扫描(OCT)的数据集(NHM4模式)中重新采样测量的视网膜神经纤维层(RNFL)厚度。方法:采用Stratus OCT(快速RNFL模式,RNFL3)和RTVue-100 OCT(NHM4(RNFL1)和RNFL3.45(RNFL2)模式)对36例健康受试者和76例青光眼受试者进行成像。在RTVue-100 OCT的NHM4和RNFL3.45模式下评估了测量的可重复性(类内相关系数(ICC))。通过Bland-Altman图分析了不同RNFL测量值之间的一致性。在不同的RNFL测量值之间比较了接收器工作特性(ROC)曲线(AUC)下用于区分健康和青光眼的面积。结果:NHM4和RNFL3.45模式均显示出优异的测量重现性(ICC 0.831-0.978)。 RTVue-100 OCT的两种不同模式和Stratus OCT的RNFL厚度在所有部门(包括平均值)中都相关。 RNFL1和RNFL2数据之间没有显着差异。但是,RTVue-100 OCT在青光眼中的RNFL厚度显着大于Stratus OCT测量的厚度。由RTVue-100 OCT的两种不同模式确定的RNFL厚度非常吻合(95%的一致限-6.53至6.95微米)。所有三个RNFL测量结果均显示出良好的青光眼辨别能力(AUC = RNFL1 0.970,RNFL2 0.962,RNFL3 0.971)。结论:通过直接扫描和从RTVue-100 OCT数据集重新采样确定的RNFL厚度吻合良好。但是,青光眼受试者的两种测量结果均与Stratus OCT显着不同。当使用不同类型的OCT对患者进行随访时,应考虑这些发现。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号