Comparison of manually corrected retinal thickness measurements from multiple spectral-domain optical coherence tomography instruments

机译:比较来自多个光谱域光学相干断层扫描仪的人工校正的视网膜厚度

获取原文
获取原文并翻译 | 示例
       

摘要

Background/aims: To compare retinal thickness measurements from three different spectral domain optical coherence instruments when manual segmentation is employed to standardise retinal boundary locations. Methods: 40 eyes of 21 healthy subjects were scanned on the Cirrus HD-OCT, Topcon 3D-OCT-2000 and Heidelberg Spectralis-OCT. Raw data were imported into custom grading software (3D-OCTOR). Manual segmentation was performed on every data set, and retinal thickness values in the foveal central subfield were computed. Results: 37 eyes of 20 subjects were gradable on every machine. The average retinal thicknesses for these eyes were 236.7 μm (SD 20.1), 235.7 μm (SD 20.4) and 236.5 μm (SD 18.0) for the Cirrus, 3D-OCT-2000 and Spectralis, respectively. Comparing manual retinal thickness measurements between any two machines, the maximum difference was 18.2 μm. The mean absolute differences per eye between two machines were: 4.9 μm for Cirrus versus 3D-OCT-2000, 3.7 μm for Cirrus versus Spectralis and 4.4 μm for 3D-OCT-2000 versus Spectralis. Conclusions: When a uniform position is used to locate the outer retinal boundary, the retinal thickness measurements derived from three different spectral domain optical coherence instruments devices are virtually identical. Manual correction may allow OCT-derived thickness measurements to be compared between devices in clinical trials and clinical research.
机译:背景/目的:当采用手动分割方法标准化视网膜边界位置时,比较三种不同光谱域光学相干仪器的视网膜厚度测量结果。方法:用Cirrus HD-OCT,Topcon 3D-OCT-2000和Heidelberg Spectralis-OCT扫描21名健康受试者的40只眼睛。原始数据被导入到定制分级软件(3D-OCTOR)中。对每个数据集进行手动分割,然后计算中央凹中央子区域的视网膜厚度值。结果:每台机器上可对20位受试者的37只眼睛进行分级。对于这些眼睛,Cirrus,3D-OCT-2000和Spectralis的平均视网膜厚度分别为236.7μm(SD 20.1),235.7μm(SD 20.4)和236.5μm(SD 18.0)。比较任意两台机器之间的人工视网膜厚度测量结果,最大差值为18.2μm。两台机器之间每只眼睛的平均绝对差为:Cirrus和3D-OCT-2000分别为4.9μm,Cirrus和Spectralis分别为3.7μm和3D-OCT-2000与Spectralis相比为4.4μm。结论:当使用统一的位置来定位视网膜外边界时,从三种不同的光谱域光学相干仪器设备获得的视网膜厚度测量值实际上是相同的。手动校正可以允许在临床试验和临床研究中比较OCT衍生的厚度测量结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号