Influence of clinical usage of GT and GTX files on cyclic fatigue resistance

机译:GT和GTX文件的临床使用对循环疲劳强度的影响

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Aim: To compare static cyclic fatigue resistance of unused, unused and sterilized and clinically used conventional NiTi GT and M-Wire GTX files. Methodology: One hundred and sixty new files (80 GT and 80 GT series X) were divided into four control groups and four experimental groups (n = 20 each). Control groups were new unused files and new sterilized files. In the experimental groups, instruments were used in three (GT1 and GTX1 groups) or four molars (GT2 and GTX2 groups). Cyclic fatigue resistance was tested in stainless steel curved canals (60°, r = 3 mm). Each file was rotated until fracture (300 rpm, 2 N cm torque). Time-to-fracture (s) was registered. Mean life, eta and beta parameters of their Weibull distributions were calculated. Results: Unused and sterilized GTX files will last significantly longer than GT files with a probability of 75% and 65%, respectively; whilst mean life was significantly longer for GT than for GTX in used files with a probability of 68%. Sterilized GT files will last longer than unused files with a probability of 66%. In both brands, unused and sterilized files will last significantly longer than files used clinically with a probability higher than 98%. The probability that GT will last longer after being used in three rather than four molars was 62% (statistically significant) and 52% (not statistically significant) for GTX. Conclusions: GTX files had an extended cyclic fatigue life when compared with GT when they were unused or unused and sterilized, but GT were significantly more resistant to cyclic fatigue after clinical usage than GT series X files 3 mm from the tip. Sterilization of files enhanced the cyclic fatigue resistance of GT instruments. Clinical use of files diminished cyclic fatigue resistance.
机译:目的:比较未使用的,未使用的,已灭菌的和临床使用的常规NiTi GT和M-Wire GTX文件的静态循环疲劳强度。方法:将160个新文件(80 GT和80 GT系列X)分为四个对照组和四个实验组(每个n = 20)。控制组是新的未使用文件和新的消毒文件。在实验组中,使用了三个(GT1和GTX1组)或四个磨牙(GT2和GTX2组)的仪器。在不锈钢弯管(60°,r = 3 mm)中测试了耐循环疲劳性。每个锉刀旋转直到断裂(300 rpm,2 N cm扭矩)。记录断裂时间。计算了其威布尔分布的平均寿命,η和β参数。结果:未使用和灭菌的GTX文件的使用寿命将比GT文件长得多,分别为75%和65%。而GT的平均寿命比旧文件中的GTX长得多,概率为68%。灭菌的GT文件比未使用的文件寿命更长,概率为66%。在这两个品牌中,未使用和消毒过的文件将比临床上使用的文件持续更长的时间,可能性高于98%。 GT在三颗而不是四颗磨牙中使用后持续更长时间的可能性对于GTX为62%(统计上显着)和52%(无统计学上显着)。结论:与未使用或未使用和消毒的GT相比,GTX锉刀具有更长的循环疲劳寿命,但与3 mm尖端的GT系列X锉刀相比,GT在临床使用后对循环疲劳的抵抗力明显更高。文件的灭菌增强了GT仪器的耐循环疲劳性。临床上使用锉刀会降低循环疲劳强度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号