Effect of macro-design of immediately loaded implants on micromotion and stress distribution in surrounding bone using finite element analysis.

机译:使用有限元分析,立即设计的植入物的宏观设计对周围骨骼的微动和应力分布的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

OBJECTIVES: Macro-design influences the initial stability of implant and reduces micromotions. The aim of this study was to determine and compare micromotions and stress distribution in the bone around immediately loaded Maestro and Xive implants using finite element analysis. MATERIALS: In this experimental study, accurate, clear photos were prepared of Xive and Maestro implants 12 and 13 mm long and 4 and 3.8 mm in diameter, respectively, using a Nikon Digital Camera with a resolution 5.24-megapixels with 8x Optical Zoom and 4x Digital Zoom. After accurate measurements, 3-D models of the implants inside the lower mandible (D2) were processed in Solidworks Version 2003 environment and transferred into Ansys for finite element analysis. After loading of 500 N angled at 70 degrees from the horizontal plane, the micromotion of the implant and Von Misses stresses around the bone were measured. RESULTS: The measured micromotion in Maestro implant was 148 mum and that in Xive was 284 mum. Stress distribution in the bone surrounding Maestro implant was better than Xive, but maximum stress surrounding Xive implants (30 MPa) was lower than Maestro (33 MPa). CONCLUSION: Based on the results obtained in the present study, maximum micromotion in maestro was less than that in Xive implants. This finding can guarantee the application of maestro implants for immediate loading.
机译:目的:宏观设计影响植入物的初始稳定性并减少微动。这项研究的目的是使用有限元分析确定并比较立即加载的Maestro和Xive植入物周围骨骼的微动和应力分布。材料:在本实验研究中,使用分辨率为5.24百万像素,具有8倍光学变焦和4倍分辨率的尼康数码相机,分别为Xive和Maestro植入物分别准备了12和13毫米长,直径分别为4和3.8毫米的准确,清晰的照片。数码变焦。经过精确测量之后,在Solidworks Version 2003环境中处理了下颌骨(D2)内植入物的3-D模型,并将其转移到Ansys中进行有限元分析。在与水平面成70度角倾斜的500 N负载后,测量植入物的微动和骨周围的Von Misses应力。结果:Maestro植入物中测得的微动为148微米,Xive中为284微米。 Maestro植入物周围骨骼的应力分布比Xive更好,但Xive植入物周围的最大应力(30 MPa)低于Maestro(33 MPa)。结论:根据本研究获得的结果,艺术大师的最大微动小于Xive植入物。这一发现可以保证将大师级植入物用于立即加载。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号