Water at the ice point: a useful quasi-blackbody infrared calibration source

机译:冰点处的水:有用的准黑体红外校准源

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Water at the ice point makes a good calibration source in the 4- to 13-μm spectral region because of its high emissivity, small skin depth, and the ease with which an ice-point bath may be prepared and used. In a simple ice bath the emissivity is slightly less than unity, and we have calculated corrections that allow one to predict the apparent radiation temperature of the equivalent blackbody. We propose an alternate configuration that uses an auxiliary mirror. This configuration should provide an emissivity extremely close to unity.
机译:冰点的水具有高发射率,较小的集肤深度以及易于准备和使用冰点浴的优点,因此在4至13μm光谱范围内,冰点处的水成为良好的校准源。在简单的冰浴中,发射率略小于1,并且我们已经计算了校正值,使人们可以预测等效黑体的视在辐射温度。我们提出了使用辅助镜的替代配置。这种配置应提供非常接近于单位的发射率。