As part of an ongoing laboratory survey process improvement program, we evaluated the addition of water to liquid scintillation cocktail to improve wipe test counting efficiency. Both polar and non-polar 3H and 14C-labeled compounds were used as model contaminants. Our results support the recommendations in the literature regarding the addition of water to scintillation cocktail We found an increase in the counting efficiency of the water-soluble material as a function of water content of the cocktail, but also observed a decrease in the efficiency of detection of the non-polar compound. The offsetting effects are believed to be the result of increased solubility of the polar compounds in water and increased quench of the already solubilized non-polar compound. The finding that adding water to the cocktail brought counting efficiencies of both polar and non-polar molecules to roughly the same value is novel and allows the use of a single quench curve for each radionuclide, regardless of chemical form.
机译:作为一个正在进行的实验室调查的一部分过程改善计划,我们评估的水液体闪烁鸡尾酒来改善擦拭测试计数效率。非极性3 h和14 c-labeled化合物为模型污染物。建议在文献中有关我们添加水闪烁鸡尾酒发现的计数效率的增加水溶性物质的水鸡尾酒的内容,但也观察到减少检测的效率非极性化合物。认为是溶解度增加的结果水和极性化合物的增加已经随着极性的淬火化合物。鸡尾酒带来的计算效率大致相同的极性和非极性分子值是小说,允许使用一个单一的猝灭曲线为每个放射性核素,不管化学形式。