Automated Dicentric Aberration Scoring for Triage Dose Assessment: 60Co Gamma Ray Dose-response at Different Dose Rates.

机译:自动化具双着丝的畸变得分分流剂量评价:60 co伽马射线的剂量反应不同的剂量率。

获取原文
获取原文并翻译 | 示例
           

摘要

The objective of this study was to establish radiation dose-response calibration curves using automated dicentric scoring to support rapid and accurate cytogenetic triage dose-assessment. Blood was drawn from healthy human volunteers and exposed to Co gamma rays at several dose rates (i.e., 1.0, 0.6, and 0.1 Gy min). After radiation, the blood was placed for 2 h in a 37 °C incubator for repair. Blood was then cultured in complete media to which a mitogen (i.e., phytoghemagglutinin, concentration 4%) was added for 48 h. Colcemid was added to the culture at a final concentration of 0.2 μg mL after 24 h for the purpose of arresting first-division metaphase mitotics. Cells were harvested at the end of 48 h. Samples were processed using an automated metaphase harvester and automated microscope metaphase finder equipped with a suite of software including a specialized automated dicentric scoring application. The data obtained were used to create dose-response tables of dicentric yields. The null hypothesis that the data is Poisson-distributed could not be rejected at the significance level of α = 0.05 using results from a Shiny R Studio application (goodness-of-fit Poisson). Calibration curves based on linear-quadratic fits for Co gamma rays at the three different dose rates were generated using these data. The calibration curves were used to detect blind test cases. In conclusion, using the automated harvester and automated microscope metaphase finder with associated automated dicentric scoring software demonstrates high-throughput with suitable accuracy for triage radiation dose assessment.
机译:本研究的目的是建立辐射剂量效应校正曲线自动化支持快速、具双着丝的得分准确的细胞遗传学诊断剂量评定。血从人类志愿者和健康暴露于公司伽马射线在几个剂量率(例如,1.0,0.6,和0.1 Gy min)。辐射,血液被2 h的37岁°C孵化器进行维修。在完成媒体的有丝分裂原(例如,phytoghemagglutinin,浓度4%)补充道48 h。秋水仙胺添加到文化最终浓度为0.2μg毫升后24小时逮捕甲级联赛中期的目的有丝分裂。使用一个自动化的h。样本处理中期收割机和自动显微镜中期仪配备一套软件包括一个专门的自动化具双着丝的评分程序。被用来创建剂量反应表的具双着丝的产量。数据是Poisson-distributed无法拒绝在显著性水平α= 0.05结果从一个闪亮的R工作室程序(拟合优度泊松)。基于线性二次适合公司伽马射线在生成的三个不同的剂量率使用这些数据。用于检测盲目的测试用例。使用自动化的收割机和自动化显微镜与相关中期仪具双着丝的自动评分软件演示高通量与适合分类的准确性辐射剂量评估。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号