首页> 外文期刊>Health Physics: Official Journal of the Health Physics Society >University of California Replacement of Cesium Irradiators with Alternative Technologies
【24h】

University of California Replacement of Cesium Irradiators with Alternative Technologies

机译:加州大学更换铯辐照器与替代技术

获取原文
获取原文并翻译 | 示例
           

摘要

The University of California possesses a large number of Cs-137 irradiators that are used in a wide variety of medical and research applications. The university president made a system-wide decision to reduce the potential threat of malevolent use of Cs-137 by switching wherever feasible to x-ray irradiators over a 3-y period of time. A Radioactive Source Replacement Working Group of involved faculty was formed to study the topic and to make recommendations as to when alternative technologies could offer equivalency. The Working Group concluded that x-ray irradiators could replace Cs-137 irradiators in most applications, with some likely exceptions. They found that the depth dose curve for the 320 kVp x-ray irradiator was found to be nearly identical to that of Cs-137 down to a depth in tissue of 4 cm. It was concluded that x rays (energies <= 320 keV) are more biologically effective than Cs-137 gamma rays, suggesting that lower doses of x rays will be required to achieve the same biological endpoint as Cs-137 gamma rays. A simple conversion factor for equating x-ray effects to Cs-137 effects was not recommended because relative biological effectiveness depends on multiple factors. They concluded that each experiment should be individually calibrated when converting from Cs-137 irradiators to x-ray irradiators. The lessons learned from implementing the project to date have shown the importance of having senior management buy-in, involving the research community in the decision making process and allowing for exceptions where equivalency of Cs-137 to x ray cannot be established.
机译:加州大学拥有一个大的数量的cs - 137用于辐照器各种各样的医学和研究应用程序。系统范围的减少潜在的决定威胁的恶意使用cs - 137通过切换任何可行的x射线辐照器仍一段时间。参与学院成立的工作小组研究主题和提出建议当可以提供替代技术等价性问题。x射线辐照器可以替代cs - 137辐照器在大多数应用程序中,有一些可能的例外。320千伏峰值曲线x射线辐照器被发现是几乎相同的cs - 137了4厘米的深度在组织。x射线(< = 320 keV能量)比cs - 137伽马射线,生物有效表明低剂量x射线需要达到同样的生物端点cs - 137伽玛射线。将x射线效果等同于cs - 137效果不推荐,因为相对生物有效性取决于多种因素。得出的结论是,每个实验当从单独校准cs - 137辐照器x射线辐照器。实施项目的经验教训日期显示在高级的重要性管理支持,包括研究社区在决策过程允许例外的等价cs - 137 x射线不能成立。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号