Ion losses due to initial recombination, volume recombination, and back diffusion were each determined by measurements and calculations for different size cylindrical ionization chambers and spherical ionization chambers. By measuring signal currents from these ionization chambers irradiated with (60)Co gamma rays, two groups of ion losses were obtained. (Group 1) Ion loss due to initial recombination and diffusion, which changes proportionally to the inverse of the voltage applied to the ionization chambers; (and group 2) ion loss due to volume recombination, which changes proportionally to the inverse of the square of the applied voltage. The diffusion loss was obtained separately by computing electric field distributions in the ionization chambers. It was found that diffusion loss is larger than initial recombination loss for the cylindrical ionization chambers and vise versa for the spherical ionization chambers.
机译:由于不同尺寸的圆柱形电离室和球形电离室的测量和计算,分别确定了由于初始重组,体积重组和反向扩散导致的离子损失。通过测量来自这些被(60)Coγ射线辐照的电离室的信号电流,获得了两组离子损失。 (第1组)由于初始重组和扩散而导致的离子损失,该损失与施加到电离室的电压的倒数成比例; (和第2组)由于体积复合而产生的离子损失,它与施加电压的平方的平方成反比。通过计算电离室中的电场分布分别获得扩散损耗。已经发现,对于圆柱形电离室,扩散损失大于初始复合损失,而对于球形电离室,扩散损失反之亦然。