The radiostrontium content in environmental samples was determined by chemical analysis by means of the fuming nitric acid method and ion exchange method with low-level beta counting and the newly developed method using crown-ether compound. Counting was performed with a low-background counter and a liquid scintillation counter together; the latter was the Cerenkov counting method. All results obtained by these three methods were in good agreement. The time for chemical separation of radiostrontium, using crown-ether compound, is much faster than fuming nitric acid and ion exchange methods. However, due to the high background of the liquid scintillation counter, the detection limit for Cerenkov counting is about two times higher than that for low-background counting. # 1998 Elsevier Science Ltd. All rights reserved
机译:通过发烟硝酸法,离子交换法(低含量β计数)和新开发的冠醚化合物法,通过化学分析测定环境样品中的放射性锶含量。用低本底计数器和液体闪烁计数器一起进行计数。后者是切伦科夫计数法。通过这三种方法获得的所有结果均吻合良好。使用冠醚化合物化学分离of的时间比发烟硝酸和离子交换方法要快得多。但是,由于液体闪烁计数器的高本底,切伦科夫计数的检测极限比低本底计数的检测极限高大约两倍。 #1998 Elsevier Science Ltd.保留所有权利