Safety products have been developed to help reduce the incidence of trauma caused by headgear. Previous studies have reported the characteristics of breakaway-type headgear release mechanisms with axial force application. Not all accidental releases are triggered by an axial force and it is necessary to understand the characteristics of these mechanisms with nonaxial force application. Thirteen headgear release mechanisms were tested as part of a complete headgear system. With the system attached to a plaster head and neck model a tensile force was applied to the system at 30 degrees to the sagittal plane at 2 rates. The force of activation at release and the distance traveled were determined and analyzed statistically. Force values ranged from 4.6 to 36.7 pounds and face bow travel before release ranged from 0.97 to 3.42 inches. No consistent pattern of rate dependence was observed. Several devices demonstrated the desirable combination of low force and face bow travel at release.
机译:已经开发出安全产品来帮助减少由头饰引起的外伤的发生率。先前的研究已经报道了具有轴向力施加的分离型头带释放机构的特征。并非所有意外释放都是由轴向力触发的,因此有必要了解施加非轴向力时这些机构的特性。作为完整头盔系统的一部分,测试了十三种头盔释放机制。将系统连接到石膏头颈模型上时,以2个速率在30度与矢状面之间的方向向系统施加拉力。确定释放时的激活力和行进距离,并进行统计分析。力值范围为4.6至36.7磅,释放前的脸弓移动范围为0.97至3.42英寸。没有观察到速率依赖性的一致模式。几种装置证明了释放时低力和面部弓形行程的理想组合。