The subcutaneous-implantable cardioverter-defibrillator (S-ICD) and its electrode were developed to avoid long-term complications of transvenous leads in the vasculature. We report a case of unexpected, inappropriate S-ICD shocks due to oversensing of high-amplitude, nonphysiologic, electrical noise artifacts that were not preceded by high-impedance alerts or sensing electrogram noise detections. Following explant, high-magnification X-ray imaging of the S-ICD electrode demonstrated partial fracture of the distal sensing conductor located near a short radius bend in the electrode at the electrode-header interface. Clinicians should be aware of a potential for fatigue failure fracture of the S-ICD electrode. Recommendations for systematic S-ICD follow-up and troubleshooting are discussed.
机译:皮下植入式心律转复除颤器(S-ICD)及其电极的开发是为了避免血管系统中经静脉导线的长期并发症。我们报告了一例意外的、不适当的S-ICD电击,原因是高振幅、非生理、电噪声伪影的过度感知,之前没有高阻抗警报或感应电图噪声检测。移植后,S-ICD电极的高倍X射线成像显示,远端传感导体部分断裂,位于电极头接口处的电极短半径弯曲附近。临床医生应意识到S-ICD电极可能出现疲劳失效断裂。讨论了系统性S-ICD随访和故障排除的建议。