Detection of miR-155 Using Two Types of Electrochemical Approaches

机译:使用两种电化学方法检测miR-155

获取原文
获取原文并翻译 | 示例

摘要

We present electrochemical biosensors for the analysis of a target miR-155 that work through the formation of miRNA-DNA hybrid duplexes by hybridization of miR-155 with its complementary DNA and through the subsequent intercalation of appropriate intercalating agents on gold electrodes. In the first method, the electrochemical behavior of methylene blue intercalated to the miRNA-DNA hybrid duplex, formed between miR-155 and its complementary DNA, on the gold electrode was measured. In the second method, the miRNA-DNA hybrid duplex formed between ferrocene-labeled DNA and miR-155 in solution is intercalated to the pyrenes immobilized on the electrode surface. This lead to an electrochemical signal being generated from the oxidized ferrocene accumulated on the electrode surface. For each method, the limit of detections for the detection of miR-155 were determined to be 0.96 and 2.22 pM, respectively. In addition, the developed electrochemical biosensors were effective in the quantitation of miR-155 in serum sample.
机译:我们提出了用于分析靶miR-155的电化学生物传感器,该传感器通过miR-155与其互补DNA杂交形成miRNA-DNA杂交双链体,并通过随后在金电极上插入适当的插层剂来工作。在第一种方法中,测量了在金电极上嵌入miR-155与其互补DNA之间形成的miRNA-DNA杂交双链的亚甲基蓝的电化学行为。在第二种方法中,将二茂铁标记的DNA和溶液中的miR-155之间形成的miRNA-DNA杂交双链插入固定在电极表面的芘中。这会导致电极表面上积聚的氧化二茂铁产生电化学信号。对于每种方法,检测miR-155的检测限分别为0.96和2.22 pM。此外,所开发的电化学生物传感器可有效定量血清样品中的miR-155。