A technique has been developed to simultaneously measure picogram per gram concentrations of Ba and Pb by isotope dilution mass spectrometry, as well as Pb isotopic ratios in polar ice by thermal ionisation mass spectrometry. BapO(2)(+) and Pb+ ions were employed for these determinations. A calibrated mixture of enriched Pb-205 and Ba-137 was added to the samples providing an accuracy of better than approximately 2% for Pb/Ba element ratio determinations. Interference by molecular ions in the Pb mass spectrum occurred only at Pb-204/Pb-205, but these contributions were negligible in terms of precisions expected on picogram-sized Pb samples. The technique is illustrated with measurements on Greenland firn, using a drill-core section that includes the Laki volcanic eruption of 1783-1784. The data show deviations from the element concentrations indicating volatile metal enrichments, but the Pb isotopic signature of the Laki lava could not be identified.
机译:已经开发了一种技术以通过同位素稀释质谱法同时测量每克BA和Pb的PicoICoIch,以及通过热电离质谱法在极性冰中的Pb同位素比。 使用Bapo(2)(+)和Pb +离子用于这些测定。 将富含PB-205和Ba-137的校准混合物加入到样品中,提供比Pb / BA元素比例的约2%的精度。 PB质谱中的分子离子的干扰仅发生在PB-204 / PB-205,但在皮科大小的PB样品上预期的精确方面,这些贡献可以忽略不计。 使用钻孔核心部分,通过包括Laki火山喷发为1783-1784的钻孔核心部分,通过测量来说明该技术。 数据显示与表明挥发性金属富集的元素浓度的偏差,但不能识别Laki熔岩的PB同位素特征。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号