Auditing of Monitoring and Respiratory Support Equipment in a Level III-C Neonatal Intensive Care Unit

机译:审计III-C水平新生儿重症监护病房监测和呼吸支持设备

获取原文
获取原文并翻译 | 示例
           

摘要

Background. Random safety audits (RSAs) are a safety tool but have not been widely used in hospitals. Objectives. To determine the frequency of proper use of equipment safety mechanisms in relation to monitoring and mechanical ventilation by performing RSAs. The study also determined whether factors related to the patient, time period, or characteristics of the area of admission influenced how the device safety systems were used. Methods. A prospective observational study was conducted in a level III-C Neonatal Intensive Care Unit (NICU) during 2012. 87 days were randomly selected. Appropriate overall use was denned when all evaluated variables were correctly programmed in the audited device. Results. A total of 383 monitor and ventilator audits were performed. The Kappa coefficient of interobserver agreement was 0.93. The rate of appropriate overall use of the monitors and respiratory support equipment was 33.68%. Significant differences were found with improved usage during weekends, OR 1.85 (1.12-3.06, p = 0.01), and during the late shift (3 pm to 10 pm), OR 1.59 (1.03-2.4, p = 0.03). Conclusions. Equipment safety systems of monitors and ventilators are not properly used. To improve patient safety, we should identify which alarms are really needed and where the difficulties lie for the correct alarm programming.
机译:背景。随机安全审核(RSAS)是一种安全工具,但尚未广泛用于医院。目标。通过执行RSA来确定与监测和机械通气有关的适当使用设备安全机制的频率。该研究还确定了与患者,时间段或入学领域的特征有关的因素是否影响了如何使用设备安全系统。方法。在2012年,在III-C水平的新生儿重症监护病房(NICU)中进行了预期观察研究。随机选择87天。当所有评估的变量正确编程在审计的设备中时,欺骗适当的整体使用。结果。共有383个监控和呼吸机审核。 Interobserver协议的Kappa系数为0.93。适当的整体使用监视器和呼吸支持设备的速度为33.68%。在周末期间使用的改善,或1.85(1.12-3.06,p = 0.01),以及在晚班(3pm至10 pm),或1.59(1.03-2.4,p = 0.03)期间,在1.85(1.12-3.06,p = 0.01)中,或1.59(1.03-2.4,p = 0.03)。结论。没有适当使用监视器和呼吸机的设备安全系统。为了提高患者的安全,我们应该确定真正需要的警报以及困难所针对正确的报警编程。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号