The risk of prion protein cross-infection has focussed attention on the potential hazards of protein contamination of re-usable medical devices. This study determined the frequency of protein contamination of ProSeal laryngeal mask airways (PLMA) after two cleaning procedures and tested the hypothesis that the combination of hand- and machine-washing removes protein contamination more effectively than hand-washing alone. After clinical use fifty-four PLMAs were randomly allocated to be washed by hand or by hand then machine. All PLMAs were then autoclaved at 134 degrees C for 4 minutes. After processing, each PLMA was immersed in a 1.2% solution of erythrosin B and examined for uptake of stain. The site (outer surface, bowl and edges of the cuff airway and drain tube, finger strap) and severity (nil/mild/moderate/severe) of staining was scored by a blinded observer. There were no differences in the site or severity of staining between the two cleaning procedures. Staining was detected on 89% of PLMAs that were hand-washed and 78% of PLMAs that were hand-, then machine-washed (P=0.27). When staining occurred, it was mild in 98%, moderate in 2% and was never severe. Staining was more frequent on the edge than at any other location (all comparisons: P < or = 0.01). The strap never had any staining. We conclude that residual contamination of PLMAs with protein deposits is common even when machine-washing is used to augment hand-washing before autoclaving. The infection risk associated with these deposits remains to be determined.
机译:ion病毒蛋白质交叉感染的风险已将注意力集中在可重复使用的医疗设备的蛋白质污染的潜在危害上。这项研究确定了两次清洗程序后ProSeal喉罩气道(PLMA)蛋白质污染的频率,并检验了以下假设:手洗和机洗相结合比单独洗手更有效地去除蛋白质污染。临床使用后,随机分配了54个PLMA进行手工清洗或手动清洗。然后将所有PLMA在134摄氏度下高压灭菌4分钟。加工后,将每个PLMA浸入1.2%的赤藓红B溶液中,检查污渍的吸收情况。由不知情的观察者对染色的部位(外表面,袖口气道和排泄管的碗和边缘,手指带)和染色的严重程度(零/轻度/中度/重度)进行评分。两种清洁程序之间的染色部位或严重程度没有差异。在89%的手洗PLMA和78%的手洗,然后机洗的PLMA上检测到染色(P = 0.27)。发生染色时,其轻度为98%,中度为2%,从不严重。边缘上的染色比其他任何地方都更频繁(所有比较:P <或= 0.01)。皮带从未被弄脏。我们得出的结论是,即使在高压灭菌之前使用机洗来增强手洗的情况下,蛋白质沉积物对PLMA的残留污染也是很常见的。与这些沉积物相关的感染风险仍有待确定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号