Movable gamma camera, 2020tc Imager, was light miniaturized using Si photodiode as a semiconductor sensor instead of photomultiplier tubes. To validate performance of this new camera in cardiac pool scintigraphy, multigated equilibrium radionuclide ventriculography using the moving cardiac phantom was performed with 2020tc Imager and a conventional Agner-type gamma camera (PRISM 3000). METHOD: Both measured cardiac functional values were compared with the set-up ones for the phantom. Five-, 7.5-, and 10-minute-multigated data were acquired using both cameras under constant contractile condition. Constant 5-minute-multigated data acquisitions using 2020tc Imager were also carried out with varied contractile conditions. RESULTS: Measured ejection fraction (EF) derived from 2020tc Imager and an Anger-type camera were 68.5 +/- 0.6 and 70.3 +/- 1.4%, respectively. Both of these values were absolutely close to the set-up EF value of 70%. Both of end-diastolic volume and EF showed excellent correlation between set-up and measured values with the correlation coefficient of 0.97 and 0.99, respectively. CONCLUSION: This new movable camera could provide comparative cardiac functional values with an Anger-type camera and it can be useful to evaluate acute cardiac function in a coronary care unit.
机译:可移动的伽马相机2020TC成像器,使用Si光电二极管作为半导体传感器而不是光电倍增管的光。为了验证在心脏池中的这种新相机的性能,使用2020TC成像仪和传统的Agner型伽马摄像机(棱镜3000)进行使用移动心脏幻影的多晶平衡放射性核素脑室。方法:将测量的心功能值与用于幻像的设定值进行比较。在恒定的收缩条件下使用两种相机获得五个,7.5-和10分钟的多倍数据。使用2020TC成像仪的恒定5分钟多颗数据采集也具有不同的收缩条件。结果:从2020TC成像仪和愤怒型相机衍生的测量射血分数(EF)分别为68.5 +/- 0.6和70.3 +/- 1.4%。这些值的两者绝对接近设置的EF值为70%。结束 - 舒张型体积和EF两者在设置和测量值之间分别显示出具有0.97和0.99的相关系数之间的优异相关性。结论:这种新型可移动摄像机可以提供具有愤怒型摄像机的比较心功能值,可用于评估冠状动脉护理单元中的急性心功能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号